Within surface science, a quartz crystal microbalance with dissipation monitoring (QCM-D) is a type of quartz crystal microbalance (QCM) based on the ring-down technique. It is used in interfacial acoustic sensing. Its most common application is the determination of a film thickness in a liquid environment (such as the thickness of an adsorbed protein layer). It can be used to investigate further properties of the sample, most notably the layer's softness.

Method

Ring-down as a method to interrogate acoustic resonators was established in 1954.[1] In the context of the QCM, it was described by Hirao et al.[2] and Rodahl et al.[3] The active component of a QCM is a thin quartz crystal disk sandwiched between a pair of electrodes.[4] The application of an AC voltage over the electrodes causes the crystal to oscillate at its acoustic resonance frequency. When the AC voltage is turned off, the oscillation decays exponentially ("rings down"). This decay is recorded and the resonance frequency (f) and the energy dissipation factor (D) are extracted. D is defined as the loss of energy per oscillation period divided by the total energy stored in the system. D is equal to the resonance bandwidth divided by the resonance frequency. Other QCM instruments determine the bandwidth from the conductance spectra. Being a QCM, the QCM-D works in real-time, does not need labeling, and is surface-sensitive. Current QCM-D equipment enables measuring of more than 200 data points per second.

Changes in the resonance frequency (Δf) are primarily related to mass uptake or release at the sensor surface. When employed as a mass sensor, the instrument has a sensitivity of about 0.5 ng/cm2 according to the manufacturer. Changes in the dissipation factor (ΔD) are primarily related to the viscoelasticity (softness).[5] The softness, in turn, often is related to structural changes of the film adhering at the sensor surface.

Mass sensor

When operated as a mass sensor, the QCM-D is often used to study molecular adsorption/desorption and binding kinetics to various types of surfaces. In contrast to optical techniques such as surface plasmon resonance (SPR) spectroscopy, ellipsometry, or dual polarisation interferometry, the QCM determines the mass of the adsorbed film including trapped solvent. Comparison of the "acoustic thickness" as determined with the QCM and the "optical thickness" as determined by any of the optical techniques therefore allows to estimate the degree of swelling of the film in the ambient liquid.[6] The difference in dry and wet mass measured by QCM-D and MP-SPR is more significant in highly hydrated layers as can be seen in.[7][8][9]

Since the softness of the sample is affected by a large variety of parameters, the QCM-D is useful for studying molecular interactions with surfaces as well as interactions between molecules. The QCM-D is commonly used in the fields of biomaterials, cell adhesion, drug discovery, materials science, and biophysics. Other typical applications are characterizing viscoelastic films, conformational changes of deposited macromolecules, build-up of polyelectrolyte multilayers, and degradation or corrosion of films and coatings.

References

  1. "Method for Determining the Viscoelastic Properties of Dilute Polymer Solutions at Audio-Frequencies". Journal of Applied Physics 25 (10): 1312–1320. 1954. doi:10.1063/1.1721552. Bibcode1954JAP....25.1312S. 
  2. "Resonance Emat system for acoustoelastic stress measurement in sheet metals". Review of Scientific Instruments 64 (11): 3198–3205. 1993. doi:10.1063/1.1144328. Bibcode1993RScI...64.3198H. 
  3. "A simple setup to simultaneously measure the resonant frequency and the absolute dissipation factor of a quartz crystal microbalance". Review of Scientific Instruments 67 (9): 3238–3241. 1998-06-04. doi:10.1063/1.1147494. Bibcode1996RScI...67.3238R. 
  4. "Studies of Viscoelasticity with the QCM". Piezoelectric Sensors. Springer Series on Chemical Sensors and Biosensors. 5. Berlin / Heidelberg: Springer-Verlag. 2007 (published 2006-09-08). pp. 49–109. doi:10.1007/5346_024. ISBN 978-3-540-36567-9. 
  5. "Viscoelastic, mechanical, and dielectric measurements on complex samples with the quartz crystal microbalance". Physical Chemistry Chemical Physics 10 (31): 4516–4534. 2008. doi:10.1039/b803960g. PMID 18665301. Bibcode2008PCCP...10.4516J. 
  6. "Adsorption of pNIPAM layers on hydrophobic gold surfaces, measured in situ by QCM and SPR". Langmuir 19 (17): 6837–6844. 2003. doi:10.1021/la034281a. 
  7. "Effect of Molecular Architecture of PDMAEMA–POEGMA Random and Block Copolymers on Their Adsorption on Regenerated and Anionic Nanocelluloses and Evidence of Interfacial Water Expulsion". The Journal of Physical Chemistry B 119 (49): 15275–15286. 10 December 2015. doi:10.1021/acs.jpcb.5b07628. PMID 26560798. 
  8. "Triggering Protein Adsorption on Tailored Cationic Cellulose Surfaces". Biomacromolecules 15 (11): 3931–3941. 10 November 2014. doi:10.1021/bm500997s. PMID 25233035. 
  9. Emilsson (15 April 2015). "Strongly Stretched Protein Resistant Poly(ethylene glycol) Brushes Prepared by Grafting-To". ACS Applied Materials & Interfaces 7 (14): 7505–7515. doi:10.1021/acsami.5b01590. PMID 25812004.