Medicine:Custom-made medical device
Custom-made medical device | |
---|---|
MeSH | D004864 |
A custom-made medical device, commonly referred to as a custom-made device (CMD) (Canada, the European Union, the United Kingdom) or a custom device (United States), is a medical device designed and manufactured for the sole use of a particular patient. Examples of custom-made medical devices include auricular splints, dentures, orthodontic appliances, orthotics and prostheses.
Definition
There is no globally agreed definition, but a custom-made medical device can be broadly defined as a medical device that has been designed and manufactured in accordance with a prescription from an appropriately qualified person for the sole use of a particular patient to meet their specific needs. Mass-produced medical devices that have been adapted for specific patient requirements such as customised wheelchairs, hearing aids, and spectacle frames do not typically fall within the definition of a custom-made medical device.
Definitions by jurisdiction
Jurisdiction | Definition | Legislation |
---|---|---|
Australia | A medical device that: (a) is made specifically in accordance with a request by a health professional specifying the design characteristics or construction of the medical device; and (b) is intended: (i) to be used only in relation to a particular individual; or (ii) to be used by the health professional to meet special needs arising in the course of his or her practice. | Therapeutic Goods (Medical Devices) Regulations 2002[1] |
Canada | A custom-made device, as defined in the regulations, means a medical device, other than a mass-produced medical device, that (a) is manufactured in accordance with a HCP's written direction giving its design characteristics; (b) differs from medical devices generally available for sale or from a dispenser; and (c) is for the sole use of a particular patient of that professionals, or, is for use by that professional to meet special needs arising in the course of his or her practice. | Medical Devices Regulations SOR/98-282[2][3] |
European Union | Any device specifically made in accordance with a written prescription of any person authorised by national law by virtue of that person's professional qualifications which gives, under that person's responsibility, specific design characteristics, and is intended for the sole use of a particular patient exclusively to meet their individual conditions and needs. However, mass-produced devices which need to be adapted to meet the specific requirements of any professional user and devices which are mass-produced by means of industrial manufacturing processes in accordance with the written prescriptions of any authorised person shall not be considered to be custom-made devices | Regulation (EU) 2017/745[4] |
United Kingdom | Any device specifically made in accordance with a written prescription of a registered medical practitioner, or any other person authorised to write a prescription by virtue of their professional qualification which gives, under that person's responsibility, specific design characteristics, and is intended for the sole use of a particular patient exclusively to meet their individual condition or need but the following devices are not custom-made devices—(a) mass-produced devices which need to be adapted to meet the specific requirements of a professional user, and (b) devices which are mass-produced by means of an industrial manufacturing process in accordance with the written prescriptions of any authorised person | Medical Devices Regulations 2002[5] |
United States | (1) In general
The requirements of sections 360d and 360e of this title shall not apply to a device that— (A) is created or modified to comply with the order of an individual physician or dentist (or any other specially qualified person designated under regulations promulgated by the Secretary after an opportunity for an oral hearing); (B) to comply with an order described in subparagraph (A), necessarily deviates from an otherwise applicable performance standard under section 360d of this title or requirement under section 360e of this title; (C) is not generally available in the United States in finished form through labeling or advertising by the manufacturer, importer, or distributor for commercial distribution; (D) is designed to treat a unique pathology or physiological condition that no other device is domestically available to treat; (E)(i) is intended to meet the special needs of such physician or dentist (or other specially qualified person so designated) in the course of the professional practice of such physician or dentist (or other specially qualified person so designated); or (ii) is intended for use by an individual patient named in such order of such physician or dentist (or other specially qualified person so designated); (F) is assembled from components or manufactured and finished on a case-by-case basis to accommodate the unique needs of individuals described in clause (i) or (ii) of subparagraph (E); and (G) may have common, standardized design characteristics, chemical and material compositions, and manufacturing processes as commercially distributed devices. (2) Limitations Paragraph (1) shall apply to a device only if— (A) such device is for the purpose of treating a sufficiently rare condition, such that conducting clinical investigations on such device would be impractical; (B) production of such device under paragraph (1) is limited to no more than 5 units per year of a particular device type, provided that such replication otherwise complies with this section; and (C) the manufacturer of such device notifies the Secretary on an annual basis, in a manner prescribed by the Secretary, of the manufacture of such device. |
Federal Food, Drug, and Cosmetic Act[6] |
Types
Depending on the jurisdiction, custom-made medical devices can be prescribed by various healthcare professionals working within numerous medical specialties such as dentists, hearing aid dispensers, ocularists/orbital prosthetists, orthotists, medical practitioners/physicians and prosthetists. Manufacturers of custom-made medical devices include anaplastologists, audiologists, clinical dental technicians/dental prosthetists/denturists, dental assistants/dental nurses, dental technicians, dentists, ocularists/orbital prosthetists, ophthalmologists, optometrists, orthopaedic shoe fitters, orthopedic technicians, orthotists and prosthetists.[7]
Custom-made devices | Medical specialty |
---|---|
Arch bars/wafers to facilitate orthognathic surgery | Oral and maxillofacial surgery |
Auricular splints | Plastic surgery |
Bruxism splints | Restorative dentistry |
Craniofacial prostheses | Oral and maxillofacial surgery |
Dental prostheses (complete dentures, partial dentures and obturators) | Oral and maxillofacial surgery/restorative dentistry: prosthodontics |
Dental restorations (crowns, bridges, veneers, inlays and onlays) | Restorative dentistry |
Dentofacial orthopaedic/functional appliances (activator appliance, Clark twin block, Fränkel appliance) | Orthodontics |
Hearing aid inserts/moulds | Audiology/otology |
Lingual arch | Orthodontics |
Lip bumper | Orthodontics |
Mandibular advancement splints | Sleep medicine |
Mouthguards | Sports dentistry |
Ocular prostheses | Ophthalmology |
Orthodontic retainers | Orthodontics |
Orthopaedic footwear | Orthoses |
Palatal expansion appliances | Orthodontics |
Quad helix | Orthodontics |
Speech prostheses | Speech-language pathology/speech therapy |
Transpalatal arches | Orthodontics |
Legislative requirements
Australia
In Australia manufacturers of custom-made medical devices are exempt from registering with the Australian Register of Therapeutic Goods (ARTG). Manufacturers of custom-made medical devices cannot advertise such devices directly to patients and are required to:
- Notify the Therapeutic Goods Administration that they are providing custom-made medical devices.
- Comply with the ARTG exemption conditions concerning inspection and review.
- Provide appropriate documentation with devices that they manufacture and/or supply.
- Maintain records relating to the devices that they have manufactured and/or supplied in Australia for at least five years.
- Submit an annual report with details of the custom-made medical devices that they have manufactured and/or supplied to the Therapeutic Goods Administration.[8]
Canada
In Canada, custom-made medical devices are subject to Part 2 of the Medical Devices Regulations (SOR/98-282)[2] under the Food and Drugs Act. Serious adverse incidents with medical devices must be reported to Health Canada within 72 hours.[citation needed]
European Union
Custom-made devices manufactured in the European Union are subject to Regulation (EU) 2017/745 (Medical Device Regulation [EU MDR]),[4] which replaced and repealed Directive 93/42/EEC (Medical Devices Directive [MDD]).[9] Under the MDD, manufacturers of custom-made devices were required to follow the relevant Essential Requirements set out in Annex I and the procedure set out in Annex VIII.[citation needed]
The EU MDR was published on 5 April 2017, came into force on 25 May 2017 and, following a three-year transition period, was expected to replace and repeal the MDD on 26 May 2020. But on 23 April 2020, Regulation (EU) 2020/561 was adopted which deferred the full implementation of the EU MDR for one year until 26 May 2021 so that efforts could be concentrated on the response to the coronavirus disease 2019 (COVID-19) pandemic. Under the EU MDR, manufacturers of custom-made devices are required to:
- Establish, document, implement and maintain, keep up to date and continually improve a quality management system. These requirements are provided in EU MDR Article 10(9) and are aligned with certain clauses of ISO 13485, the International Organization for Standardization (ISO) quality management system requirements for the design and manufacture of medical devices.
- Comply with the relevant General Safety and Performance Requirements set out in Annex I. These obligations are comparable with the MDD Annex I Essential Requirements but are expanded and include the requirement to establish, implement, document and maintain a risk management system, the requirements of which are in alignment with ISO 14971, the ISO standard for the application of risk management to medical devices.
- The procedure set out in Annex XIII, which is comparable with MDD Annex VIII but with some enhanced requirements.
- Review and document experience gained in the post-production phase and report serious incidents and field safety corrective actions.
- Manufacturers outside the EU who are placing medical devices on the EU market are obligated to appoint a European Authorized Representative.[10]
Custom-made devices are not required to carry the CE marking.
United Kingdom
In the UK manufacturers of custom-made devices are required to register with the Medicines and Healthcare products Regulatory Agency. Until the UK left the European Union on 31 January 2020, custom-made devices were governed by the MDD, which was given effect in UK law by The Medical Devices Regulations 2002 (Statutory Instrument 2002/618 [UK MDR 2002][5]). Immediately after the UK's departure, the UK entered an 11-month implementation period (IP), during which EU law continued to apply.
In preparation for the UK's departure from the EU, the EU MDR was essentially transposed into The Medical Devices (Amendment etc.) (EU Exit) Regulations 2019, (Statutory Instrument 2019/791 [UK MDR 2019]), an amendment of the UK MDR 2002[11]) and was expected to be fully implemented on exit day. The UK MDR 2002 was further amended by The Medical Devices (Amendment etc.) (EU Exit) Regulations 2020 (Statutory Instrument 2020/1478 [UK MDR 2020][12]), which removed the provisions of the EU MDR and substituted 'exit day' for 'IP completion day'.
In Great Britain medical devices can conform to either the UK MDR 2002 (as amended) or the EU MDR until 30 June 2023. Northern Ireland remains in line with EU law under the terms of the Protocol on Ireland/Northern Ireland.[13]
Custom-made devices are not required to carry the CE marking or the UK Conformity Assessed (UKCA) marking.[citation needed]
United States
Custom devices are subject to requirements including labelling (21 CFR Part 801), reporting (21 CFR Part 803), corrections and removals (21 CFR Part 806), registration and listing (21 CFR Part 807) and quality systems regulation (21 CFR 820). Manufacturers of custom devices are obligated to submit an annual report of custom devices to the Food and Drug Administration but are exempt from Premarket Approval (PMA) requirements and conformance to mandatory performance standards.[14]
See also
- Medical device
- Canada: Marketed Health Products Directorate
- United States: Medical Device Regulation Act
References
- ↑ "Therapeutic Goods (Medical Devices) Regulations 2002 (Statutory Rules No. 236, 2002)" (in English). Regulations of 14 January 2022. Australia. https://www.legislation.gov.au/Details/F2022C00061.
- ↑ 2.0 2.1 "Medical Devices Regulations SOR/98-282" (in English). Regulations of 7 May 1998. https://laws-lois.justice.gc.ca/PDF/SOR-98-282.pdf.
- ↑ Guidance for Health Care Professionals on Special Access and Custom-Made Medical Devices. Ottawa, Ontario, Canada: Health Canada. 18 February 2016. https://www.canada.ca/en/health-canada/services/drugs-health-products/special-access/medical-devices/guidance-health-care-professionals-special-access-custom-made.html.
- ↑ 4.0 4.1 "European Parliament / European Council. Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC". Article 2, Regulation No. (EU) 2017/745 of 5 April 2017. The European Parliament and Council of the European Union. p. 16. https://eur-lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32017R0745.
- ↑ 5.0 5.1 "The Medical Devices Regulations 2002 (Statutory Instrument 2002/618)" (in English). of 20 May 2002. United Kingdom. https://www.legislation.gov.uk/uksi/2002/618/pdfs/uksi_20020618_en.pdf.
- ↑ "Federal Food, Drug, and Cosmetic Act". Section 520(b), of 11 December 2020. p. 425–426. https://www.govinfo.gov/content/pkg/COMPS-973/pdf/COMPS-973.pdf.
- ↑ "Custom-made devices in Great Britain". Medicines and Healthcare products Regulatory Agency. 31 December 2020. https://www.gov.uk/government/publications/custom-made-medical-devices/custom-made-devices-in-great-britain.
- ↑ "Custom-made medical devices: Information for sponsors, health professionals & manufacturers". Therapeutic Goods Administration. November 2021. https://www.tga.gov.au/sites/default/files/custom-made-medical-devices.pdf.
- ↑ "Council Directive 93/42/EEC of 14 June 1993 [1 concerning medical devices, OJ No L 169/1 of 1993-07-12"]. Article 2, of 14 June 1993. Council of the European Union (Medical Device Directive). https://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=CONSLEG:1993L0042:20071011:en:PDF.
- ↑ Green, James I. J. (15 March 2021). "Medical Device Regulation: Requirements for Dental Professionals Who Prescribe and Manufacture Custom-Made Devices". Primary Dental Journal 10 (1): 64–88. doi:10.1177/2050168420980980. PMID 33722134.
- ↑ "The Medical Devices (Amendment etc.) (EU Exit) Regulations 2019 (Statutory Instrument 2019/791)" (in English). of 1 April 2019. United Kingdom. https://www.legislation.gov.uk/uksi/2019/791/made/data.pdf.
- ↑ "The Medical Devices (Amendment etc.) (EU Exit) Regulations 2020 (Statutory Instrument 2020/1478)" (in English). of 8 December 2020. United Kingdom. https://www.legislation.gov.uk/uksi/2020/1478/pdfs/uksi_20201478_en.pdf.
- ↑ (in English) of 17 October 2019. United Kingdom. https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/840230/Revised_Protocol_to_the_Withdrawal_Agreement.pdf.
- ↑ Custom Device Exemption: Guidance for Industry and Food and Drug Administration Staff. Washington, D.C., United States: Federal government of the United States. 14 January 2014. https://www.fda.gov/media/89897/download.
Original source: https://en.wikipedia.org/wiki/Custom-made medical device.
Read more |