Engineering:Vector signal analyzer

From HandWiki
A vector signal analyzer display featuring a constellation diagram, demodulation error data, signal spectrum, and the real-time measured signal

A vector signal analyzer is an instrument that measures the magnitude and phase of the input signal at a single frequency within the IF bandwidth of the instrument. The primary use is to make in-channel measurements, such as error vector magnitude, code domain power, and spectral flatness, on known signals.

Vector signal analyzers are useful in measuring and demodulating digitally modulated signals like W-CDMA, LTE, and WLAN.[1] These measurements are used to determine the quality of modulation and can be used for design validation and compliance testing of electronic devices.

Operation

A vector signal analyzer block diagram featuring a down-convert stage, a digitizing stage, and a DSP and display stage

The vector signal analyzer spectrum analysis process typically has a down-convert & digitizing stage and a DSP & display stage.

Down-convert and digitize stage

A vector signal analyzer operates by first down-converting the signal spectra by using superheterodyne techniques. A portion of the input signal spectrum is down-converted (using a voltage-controlled oscillator and a mixer) to the center frequency of a band-pass filter. The use of a voltage-controlled oscillator allows consideration of different carrier frequencies.

After the conversion to an intermediate frequency, the signal is filtered in order to band-limit the signal and prevent aliasing. The signal is then digitized using an analog-to-digital converter. Sampling rate is often varied in relation to the frequency span under consideration.

DSP and display stage

Once the signal is digitized, it is separated into quadrature and in-phase components using a quadrature detector, which is typically implemented with a discrete Hilbert transform. Several measurements are made and displayed using these signal components and various DSP processes, such as the ones below.

Signal spectrum from FFT

Main page: Fast Fourier transform

A FFT is used to compute the frequency spectrum of the signal. Usually there is a windowing function option to limit spectral leakage and enhance frequency resolution.[2] This window is implemented by multiplying it with the digitized values of the sample period before computing the FFT.

Constellation diagram

Main page: Constellation diagram

A constellation diagram represents a signal modulated by a digital modulation scheme such as quadrature amplitude modulation or phase-shift keying. This diagram maps the magnitude of the quadrature and in-phase components to the vertical and horizontal directions respectively. Qualitative assessments of signal integrity can be made based on interpretation of this diagram.

Error vector magnitude

Main page: Engineering:Error vector magnitude

By representing the quadrature and in-phase components as the vertical and horizontal axes, the error vector magnitude can be computed as the distance between the ideal and measured constellation points on the diagram. This requires knowledge of the modulated signal in order to compare the received signal with the ideal signal.

Typical functionality

Typical vector signal analyzer displays feature the spectrum of the signal measured within the IF bandwidth, a constellation diagram of the demodulated signal, error vector magnitude measurements, and a time-domain plot of the signal. Many more measurement results can be displayed depending on the type of modulation being used (symbol decoding, MIMO measurements, radio frame summary, etc.).

References

Footnotes

  1. National Institute of Standards and Technology [1], Measurement to Support Modulated-Signal Radio Transmissions for the Public-Safety Sector, p. 15, April, 2008, accessed July 19, 2011.
  2. Keysight [2], Keysight Vector Signal Analysis Basics, Published September 30, 2019.