Medicine:Pulmonary diffusing capacity for nitric oxide
The pulmonary diffusing capacity for nitric oxide (DL,NO), also called the transfer factor of the lung for nitric oxide (TL,NO), is a pulmonary function measurement that quantifies the rate of uptake of inhaled nitric oxide (NO) from the alveolar gas into the pulmonary capillary blood. It is expressed in units of mL·min−1·mmHg−1 (traditional) or mmol·min−1·kPa−1 (SI), with the conversion factor being division by 2.985. Because NO reacts with haemoglobin roughly 1,500 times faster than carbon monoxide (CO), DL,NO is weighted toward the alveolar–capillary membrane component of gas transfer (DM), in contrast to DL,CO, which is dominated by the red cell or blood component (θCO·VC).[1][2] When measured simultaneously with DL,CO, the combined single-breath DL,NO–DL,CO technique allows partitioning of total pulmonary gas transfer resistance into its membrane (DM) and microvascular (θ·VC) components using the Roughton–Forster equation.[3]
The test was first performed in 1983 and has been the subject of a 2017 European Respiratory Society (ERS) technical standards document that standardised its measurement methodology, equipment specifications, quality control requirements, and reference equations for clinical use.[2]
History
Origins in Cambridge
Early research on nitric oxide in the United Kingdom was directed toward its toxicology rather than its physiology. In the late 1970s, "oxides of nitrogen" were suspected as a cause of lung disease from atmospheric pollution, indoor air pollution, and cigarette smoke. High concentrations of NO (over 100 parts per million) and lower doses of nitrogen dioxide (NO2; 25 ppm) were known to cause lung damage in cases of accidental human exposure or experimental animal exposure.[1] An emphysema-like lesion had been described in laboratory rats exposed to nitrogen oxides, fuelling speculation that these compounds caused emphysema in smokers.[1]
At that time, the tobacco industry was under pressure to produce a "safer cigarette." To investigate NO in cigarette smoke, scientists at British American Tobacco constructed a prototype NO gas analyser based on chemiluminescence, a technique described by Brian Thrush and colleagues in 1964.[4] The analyser generated ozone from atmospheric oxygen and reacted it with the gas sample; any NO was immediately converted to NO2 in an excited state, which emitted photons of light upon returning to the ground state. This highly sensitive and specific method has underpinned most subsequent work on pulmonary NO uptake.[1]
Tim Higenbottam at Cambridge collaborated with British American Tobacco, who loaned him their analyser. Higenbottam and colleagues had previously failed to demonstrate a relationship between smokers' cigarette tar yields and airflow obstruction prevalence, and concluded that attention should be paid to the gas-phase constituents of smoke, principally CO and NO.[5] Colin Borland's doctoral project was to compare NO and CO uptake using the chemiluminescent analyser for NO alongside a commercial infrared CO analyser with a helium analyser. He initially examined 40 ppm—a concentration he had (erroneously) calculated to be the alveolar NO level during inhalation of a popular UK cigarette brand. Borland found that NO uptake only occurred when volumes exceeding the dead space were inhaled.[1]
The first measurements of DL,NO (reported as TL,NO) were presented by Borland, Chamberlain and Higenbottam to the UK Medical Research Society in Spring 1983.[6] The suggestion to measure the transfer factor came from their technician, Andrew Chamberlain (subsequently Professor of Bioarchaeology at the University of Manchester).[1] They found that KNO (the rate constant for alveolar NO uptake) was highly correlated with but exceeded KCO four- to five-fold, and like KCO, increased with exercise. In a subsequent abstract, they calculated DM assuming θNO to be infinite.[7] The full paper, containing observations on varying breathhold time, undetectable back tension, greater volume dependence of DL,NO than DL,CO, and independence from hyperoxia, was published in 1989.[8]
Regarding the original hypothesis, it is now considered very unlikely that NO in cigarette smoke causes emphysema: the half-life of oxidation of NO in air is approximately one hour, compared to roughly two seconds for alveolar uptake. Furthermore, histological NO2-induced emphysema in rats differs markedly from smokers' emphysema, and large international differences in cigarette NO yields do not parallel emphysema incidence.[1]
The French connection
The Bordeaux contribution arose independently, a positive side-effect of the May 1968 events in Paris. Professor Daniel Bargeton, Hervé Guénard's mentor, was vice-dean of the Paris medical faculty. When the university came to a standstill during the social upheaval, he returned to research and conceived the idea of calculating DM,CO and VC using two transfer gases in a single manoeuvre. He initially proposed using CO and hydrogen sulphide (H2S), but this gas proved highly problematic: its odour was objectionable, it deposited sulphur in mass spectrometer ionisation chambers, and it was too soluble in water, disappearing rapidly into the airway walls.[1]
After arriving in Bordeaux in 1979, Guénard systematically reviewed a list of candidate gases from a chemistry textbook. Sulphur compounds largely shared the same drawbacks; the list of nitrogen oxides was shorter, and the choice of NO was straightforward since other oxides such as N2O2 and NO2 did not meet specifications.[1] Guénard reasoned that the equation 1/DL = 1/DM + 1/(θ·VC) could be solved by a single-breath manoeuvre using CO and NO simultaneously. The Bordeaux group published their formula for DM and VC from a combined single-breath DL,NO and DL,CO in 1987.[3] Importantly, among their references was a 1958 paper by Carlsen and Comroe who had measured the rate constant for the reaction of NO with red cells in the absence of oxygen using the same rapid-reaction apparatus that Forster had used for CO.[9]
Two other significant developments occurred in 1987: two independent groups proved that NO was identical to Endothelium-Derived Relaxing Factor (EDRF), enormously increasing scientific interest in NO;[10][11] and Forster made further measurements of θCO at physiological pH, obtaining values that yielded slightly lower VC and rather larger DM compared to his 1957 data.[12]
Evolution (1989–2016)
By the late 1980s, analysers capable of detecting NO concentrations down to 1 part per billion became available. This allowed detection of endogenous NO back tension at approximately 10 ppb, enabled longer breathhold times than 7.5 seconds, and permitted alveolar profile measurements using rapidly responding analysers for steady-state and intrabreath methods.[1] Commercial gas transfer systems incorporating NO analysers also became available, using less expensive but less sensitive NO electrochemical cells that require shorter breathhold times.
Over the subsequent 25 years, numerous studies measured combined DL,NO and DL,CO in normal volunteers and in patients with a variety of cardiopulmonary diseases. Key findings included that DL,NO increases more with lung volume than DL,CO, increases linearly with exercise, increases at altitude in newcomers within 2–3 days (returning toward baseline by 7–8 days), and is reduced during diving while breathing oxygen.[1][2]
The θNO debate: finite or infinite?
A central and ongoing debate concerns whether the specific blood conductance for NO (θNO) should be regarded as effectively infinite—meaning that DL,NO equals DM,NO directly—or finite with a meaningful blood resistance to NO transfer. The original Bordeaux assumption of infinite θNO was widely adopted by other investigators.[1]
The Cambridge group challenged this view with a series of experiments. Using a membrane oxygenator model, they demonstrated that DL,NO increased upon haemolysis, proving that blood resistance to NO transfer was significant in the oxygenator.[13] They subsequently showed near-identical increases in DL,NO with cell-free haemoglobin in both an in vivo canine model (using a bovine haemoglobin-based blood substitute) and their in vitro oxygenator, establishing that the resistance of blood to NO transfer was significant in the lung.[14] Further work localised this resistance to the red cell interior specifically, rather than plasma or the red cell membrane.[15]
The 2017 ERS Task Force established the current consensus that θNO should be taken as a finite value of 4.5 mL·min−1·mmHg−1·mL−1 of blood, as measured by Carlsen and Comroe in 1958.[2][9]
Physiological basis
Roughton–Forster equation
The uptake of both NO and CO from the alveolus to the capillary blood is described by the Roughton–Forster equation:[16]
where 1/DL is the total transfer resistance, 1/DM is the membrane resistance (from the alveolar epithelial surface to the red cell membrane), and 1/(θ·VC) is the red cell or blood resistance. DM is the alveolar–capillary membrane diffusing capacity, θ is the specific conductance of blood (the volume of gas taken up by red cells per mL of blood per minute per mmHg of partial pressure), and VC is the total volume of blood in the lung capillaries exposed to alveolar air.[2]
The chief barrier to CO uptake is within the red cell (approximately 70–80% of total resistance), with approximately 25% of resistance located in the alveolar membrane. For NO, the main resistance lies between the alveolar and red blood cell membranes (approximately 60%), with the red cell resistance comprising approximately 40% of the total.[2][14]
Reaction of NO and CO with capillary blood
The reaction of haemoglobin in solution with NO is extremely rapid—nearly 1,500 times faster than with CO.[17] Crucially, the reaction of NO with haemoglobin solutions is 500–1,000 times faster than its reaction with intact erythrocytes from animal or human sources, demonstrating that θNO cannot be "infinite."[2] This red cell resistance was localised to the cell interior by experiments progressively converting oxyhaemoglobin to methaemoglobin by adding nitrite, which altered DL,NO; by contrast, altering extracellular viscosity or comparing different erythrocyte sizes did not affect DL,NO.[15]
Alveolar–capillary membrane diffusing capacity and the α-ratio
The membrane diffusing capacity DM encompasses the entire diffusion path from the surfactant lining layer through the alveolar epithelium, interstitium, capillary endothelium and plasma to the haemoglobin molecule. Its determinants are tissue diffusivity and the pressure gradient between alveolus and plasma. For a gas in tissue, diffusivity is the ratio of solubility to the square root of molecular weight. NO and CO have similar molecular weights (30 and 28 g·mol−1), but NO has approximately twice the solubility of CO in tissue. The resulting diffusivity ratio DM,NO/DM,CO is generally taken as α = 1.97.[18][2]
Thus, DM,NO = α · DM,CO = 1.97 · DM,CO. This relationship, together with a finite θNO and an appropriate θCO equation, allows DM,CO and VC to be calculated from simultaneously measured DL,NO and DL,CO in a single manoeuvre.[2]
NO in the gas phase
Airway uptake of inhaled NO in a single breathhold is negligible (approximately 0.02%). Within the acinus, the dominant mode of gas transport is molecular diffusion. Gas-phase diffusion coefficients are inversely proportional to the square root of molecular weight, so there is no significant difference between NO and CO. Gas-phase diffusion resistance as a proportion of total transfer resistance is greater for NO than for CO, but in normal lungs the effect is negligible (approximately 5% of total 1/DL,NO).[2]
Diffusion-limited uptake
Like CO, the uptake of NO is diffusion-limited rather than perfusion-limited. This is demonstrated by a very low DL/(β·Q̇) ratio of approximately 0.012 at rest (where β is the capacitance coefficient and Q̇ is cardiac output), indicating that the diffusive conductance is the rate-limiting step in alveolar NO uptake. With exercise, this ratio decreases further since the increase in cardiac output substantially exceeds the increase in DL,NO. A constant DL,NO with 25-fold variation in blood flow in a membrane oxygenator model also supports diffusion rather than perfusion limitation.[2]
Time-based diffusion–reaction framework
An alternative to the classic Roughton–Forster resistance model was developed by Kang, Sapoval and colleagues, who solved the diffusion–reaction equations for simplified capillary–red cell geometries from first principles rather than treating the uptake process as a sum of independent resistances.[19] In this framework, gas uptake is described in terms of characteristic times—an equivalent diffusion time for transport from the alveolar surface to the red cell membrane, and a reaction time for the intracellular haemoglobin reaction—rather than as a reciprocal sum of membrane and blood conductances. This approach was subsequently extended into a comprehensive model of acinar gas exchange during both breath holding and tidal breathing.[20]
A central finding is that CO and NO exhibit fundamentally different capture mechanisms within the red cell. The penetration depth of CO—determined by the ratio of intracellular diffusivity to the haemoglobin reaction rate—is approximately 1 μm, comparable to the red cell half-thickness. Consequently, CO diffuses throughout the entire red cell volume before being bound, and DL,CO behaves as a volume absorption process that scales with haematocrit and red cell volume.[19] For NO, the haemoglobin reaction is so rapid that the penetration depth is only a small fraction of the red cell radius. Incoming NO molecules are captured almost immediately upon crossing the red cell membrane, and the interior of the cell contributes negligibly to uptake. DL,NO therefore behaves as a surface absorption process, dominated by the accessible red cell surface area and the morphology of the membrane–plasma path between the alveolar wall and the red cell surfaces.[19][20]
This framework also demonstrated that the two resistance terms in the Roughton–Forster equation are not fully independent: changing the membrane thickness in computational models altered the derived blood resistance, violating the assumption of strict series additivity on which back-calculation of DM and VC from the classic equation depends.[19] Additionally, the conventional extrapolation of 1/DL,CO to zero alveolar PO2—originally proposed to isolate the membrane component—was shown not to yield the true diffusive conductance, potentially introducing error into VC estimates obtained by this method.[19]
Clinically, the time-based interpretation supports treating DL,NO and DL,CO as two complementary but independent measurements rather than solely as inputs to a component-splitting equation. DL,NO provides information about the alveolar membrane–plasma architecture and accessible red cell surface, while DL,CO reflects red cell volume and haematocrit. Both measurements, together with VA and the DL,NO/DL,CO ratio, can be reported as primary observables, with any DM and VC values flagged as model-dependent.[20][21]
Determinants
Lung volume
In normal subjects, DL,NO decreases to a greater extent than DL,CO when lung volume declines. Compared to 100% of alveolar volume (VA), DL,NO decreases by approximately 40% when VA is halved, whereas DL,CO only decreases by approximately 25% for the same reduction. Consequently, for the same decrease in lung volume, the percentage increase in KCO (DL,CO/VA) is approximately double that of KNO (DL,NO/VA), reflecting the greater dependence of DL,NO on the membrane-to-volume ratio (DM,NO/VA) than on the capillary volume-to-volume ratio (VC/VA).[2]
Blood flow and exercise
DL,NO increases linearly with exercise intensity. In healthy subjects, there is a linear increase in DL,NO of approximately 16–22 mL·min−1·mmHg−1 for every 1.0 L·min−1 increase in oxygen uptake, or approximately 5–7 mL·min−1·mmHg−1 for every 1.0 L·min−1 increase in cardiac output.[2] Notably, DL,CO correlates more tightly with cardiac output than DL,NO, suggesting that DL,CO is more sensitive than DL,NO to alveolar microvascular recruitment.[2] Both DL,NO and DL,CO are impaired for several hours following strenuous exercise, likely due to mild interstitial pulmonary oedema and/or reduced pulmonary capillary blood volume.[22]
Posture
After adjusting for postural changes in VA, both DL,NO and DL,CO increase approximately 5% from the upright sitting to the supine position, attributable to a roughly 13% increase in VC when supine.[2]
Altitude and diving
After 2–3 days at high altitude (4,400–5,000 m), both DL,NO and DL,CO increase at rest in healthy lowlanders. The DL,NO/DL,CO ratio falls acutely (by approximately 8%), returning toward baseline after a week. These changes are explained by alveolar expansion (weighted by DL,NO) and capillary recruitment (weighted by DL,CO) secondary to hyperventilation and increased cardiac output.[2]
Diving has biphasic effects: both DL,CO and DL,NO increase transiently after short compressed-air or maximal breathhold dives due to pulmonary vasodilation and central blood volume shifts, followed by parallel decreases reflecting interstitial oedema and ventilation–perfusion mismatch. Longer-duration dives are associated with reduced DL,CO owing to oxygen toxicity.[2]
Back tension
Endogenous alveolar NO concentration is approximately 8–20 ppb during tidal breathing and approximately 100–140 ppb in the nose. Using inhaled NO concentrations of 40–60 ppm with a nose clip effectively eliminates back-tension interference. The presence of endogenous NO does not measurably affect DL,NO or DL,CO.[2]
Heterogeneity
A drawback of the single-breath DL,NO measurement is that the collected alveolar sample (500–1,000 mL) is not truly representative of the actual dispersion of function within even normal lungs. Ventilatory heterogeneity leads to underestimation of both DL,NO and DL,CO compared to the homogeneous situation. DL,NO and DL,CO both decrease as breathhold time is prolonged, because the decrease in KNO and KCO at longer breathhold times (giving more weight to slowly-filling compartments) outweighs the increase in accessible alveolar volume. Since DL,NO and DL,CO are similarly affected, the effect of heterogeneity on the DL,NO/DL,CO ratio is small in normal subjects.[2]
Clinical applications
The DL,NO/DL,CO ratio
As DL,NO is weighted by DM and DL,CO is weighted by VC, the DL,NO/DL,CO ratio reflects the relative change in membrane-to-capillary components of uptake (DM,CO/VC). An increase in the DL,NO/DL,CO ratio signifies a reduction in VC that outweighs any reduction in DM—that is, predominantly microvascular pathology. A decrease in the ratio suggests greater disruption of alveolar membrane structures relative to microvascular change. Since DL,NO is relatively insensitive to haematocrit changes, the ratio should also rise in anaemia.[1][2] The normal mean DL,NO/DL,CO ratio is 4.79 ± 0.40.[2]
Microvascular disease
In pulmonary arterial hypertension (PAH), studies have shown a predominantly microvascular pattern, with a reduction in VC greater than the reduction in DM,CO, leading to elevated DM,CO/VC and DL,NO/DL,CO ratios.[2] In liver cirrhosis with hepatopulmonary syndrome, a similar rise in the DL,NO/DL,CO ratio consistent with microvascular disease has been reported.[2]
Interstitial lung disease
Results in interstitial lung disease have been variable, likely reflecting differences in pathophysiology and clinical stage. A greater reduction in DM,CO than VC (with a fall in the DL,NO/DL,CO ratio) was observed in sarcoidosis using a rebreathing technique, whereas the opposite pattern was found in diffuse parenchymal lung disease with PAH using a single-breath technique.[2]
Airflow obstruction
In asymptomatic smokers without airflow obstruction (GOLD stage 0), DM,CO was preserved relative to VC, and the DL,NO/DL,CO and DM,CO/VC ratios were elevated compared to controls, suggesting that a reduction in VC may be an early sign of chronic obstructive pulmonary disease (COPD). In established COPD, both DM and VC appear to be reduced.[2]
Miscellaneous conditions
Reduced DL,NO and DL,NO/DL,CO ratios have been reported in chronic renal failure (after adjusting for haemoglobin). In morbid obesity, there is slight reduction in DM,CO/VC. In cystic fibrosis, both DM,CO/VC and DL,NO/DL,CO are reduced. Following bone marrow transplant, both DL,NO and DL,CO are reduced.[2]
Enhanced disease detection: the double diffusion advantage
A growing body of evidence from individual-participant data meta-analyses demonstrates that combining DL,NO with DL,CO improves the classification and prediction of cardiopulmonary disease compared with DL,CO alone or either measure in isolation.
Heart failure
In a retrospective analysis of 140 patients with New York Heart Association Class II heart failure (ejection fraction <40%) and 50 controls, Zavorsky and Agostoni evaluated whether DL,NO added predictive value to DL,CO testing.[23] Of 12 models evaluated using the Bayesian information criterion (BIC), the top two models were combined DL,NO + DL,CO z-score models. The highest Matthews correlation coefficient (MCC = 0.51) was achieved by combined z-score models, indicating moderate classification ability. Combined z-scores explained 32% of the variation in the likelihood of having heart failure, which exceeded the explanatory power of either DL,NO or DL,CO z-scores alone. While DL,CO reflects predominantly pulmonary vascular and blood volume abnormalities (approximately 70–80% of the red blood cell barrier), DL,NO captures the diffusion path between the alveolar–capillary membrane and the red blood cell membrane (approximately 60%); together, the two measurements sample complementary diffusion pathways.[23]
Post-COVID-19 lung impairment
In an individual-participant data meta-analysis pooling data from 256 COVID-19 survivors and 76 controls across six centres in Italy, France, Spain and Australia, Zavorsky et al. assessed whether combining DL,NO and DL,CO enhanced detection of COVID-19-related lung pathology compared with DL,CO alone or other standard pulmonary function measures.[24] Among 256 subjects with complete pulmonary function testing post-COVID-19 (median age 60 years), 57% showed some form of impairment (airway obstruction, restriction, mixed disorder, or DL,NO or DL,CO below the lower limit of normal). Of 34 models evaluated, the model with the lowest BIC was a combined DL,NO + DL,CO z-score model, demonstrating superior COVID-19 detection. Five of the top nine models were combined DL,NO + DL,CO z-score models, and MCC values indicated that six of the top nine models were either combined or DL,NO-only models. Dyspnoea severity correlated with combined z-scores (p < 0.001).[24]
Emphysema in smokers
In a multicentre individual-participant data meta-analysis of 408 adult smokers (85 with CT-defined emphysema, 323 without) from three European hospital centres using a standardised 10-second breathhold double diffusion protocol, Zavorsky et al. evaluated whether DL,NO z-scores outperform DL,CO z-scores for emphysema detection.[25] Among 34 candidate logistic models, the lowest BIC (164.6) occurred for a parsimonious three-predictor model comprising total lung capacity (TLC), forced expiratory volume in one second (FEV1), and DL,NO z-scores (Model C), with an 88% probability of being superior to the next-lowest BIC model. Discrimination was excellent (area under the receiver operating characteristic curve 0.97, 95% CI 0.95–0.98) and classification was high (MCC 0.80, 95% CI 0.69–0.89). Hierarchical partitioning showed unique contributions from FEV1 z-scores (R2 = 0.35) > DL,NO z-scores (R2 = 0.21) > TLC z-scores (R2 = 0.11), totalling a McFadden's R2 of 0.66. Critically, adding DL,CO z-scores to the DL,NO-based model increased the total R2 trivially (by 0.003) and worsened BIC, indicating that the unique information captured by DL,CO did not add a meaningful predictive signal for emphysema beyond DL,NO. This finding is physiologically plausible: emphysema produces early alveolar–capillary membrane destruction, the compartment to which DL,NO is preferentially sensitive.[25]
Summary
Across these three disease contexts—heart failure, post-COVID-19 lung impairment and emphysema—DL,NO consistently contributed independent predictive information beyond that provided by DL,CO, spirometry or lung volumes. These findings support the recommendation that, whenever a DL,CO test is clinically indicated and the necessary equipment is available, DL,NO should be measured in the same manoeuvre and both values reported together with their ratio.[21]
ERS technical standards
In 2017, a panel of experienced physicians, physiologists and a technologist, funded by the European Respiratory Society, published a consensus document standardising the technique and application of single-breath DL,NO measurement.[2]
Gas analysers and equipment
All commercially available DL,NO apparatus is based on the single-breath DL,CO system with the addition of NO in the inspiratory gas mixture and an NO analyser. Because NO reacts with oxygen to form NO2 at a rate of approximately 1.2 ppm·min−1 (at 60 ppm NO in 21% O2), NO gas is stored separately in a high-concentration cylinder with nitrogen (typically 400–1,200 ppm NO in N2) and should be mixed with the remaining inspiratory gases no more than two minutes before use.[2]
Two types of NO analysers are available. Chemiluminescence analysers are highly sensitive (detection limit 0.5 ppb, linear to 500 ppm, response time approximately 70 ms) but expensive. The less expensive NO electrochemical cell, used in most commercial pulmonary function systems, has a detection range of 0–100 ppm and a response time below 10 seconds. The lower sensitivity of the electrochemical cell necessitates a shorter breathhold time.[2]
Testing technique
The DL,NO test follows the general methodology of the single-breath DL,CO manoeuvre. Subjects should refrain from smoking for 12 hours and from vigorous exercise for 12 hours prior to testing. After a period of quiet tidal breathing, the subject performs a rapid inspiration from residual volume to total lung capacity (achieving ≥90% of inspiratory vital capacity in less than 2.5 seconds) of a test gas mixture containing NO (40–60 ppm), CO (approximately 0.3%), an inert tracer gas (helium, methane or neon), and close to 21% oxygen. The breathhold time is 10 seconds when using a chemiluminescence analyser, or 4–6 seconds with an electrochemical cell. Following breathhold, the subject exhales smoothly and rapidly to residual volume.[2]
An initial washout volume (0.75–1.0 L) of dead space gas is discarded, and a 0.5–1.0 L alveolar sample is then collected for analysis. Between successive tests, an interval of at least 4–5 minutes is allowed for elimination of prior test gases. The effective breathhold time is calculated using the Jones–Meade formula.[26] The expired alveolar oxygen concentration should be measured so that θCO can be calculated from the alveolar PO2.[2]
Calculations
The following consensus values are used for calculating DM,CO and VC from the simultaneous one-step NO–CO technique:[2]
| Parameter | Consensus value |
|---|---|
| θNO | 4.5 mL NO·(mL blood·min·mmHg)−1 |
| 1/θCO | (0.0062 · PAO2 + 1.16) · (ideal Hb ÷ measured Hb) |
| Ideal Hb (males) | 14.6 g·dL−1 |
| Ideal Hb (females) | 13.4 g·dL−1 |
| DM,NO/DM,CO ratio (α) | 1.97 |
| Breathhold time | 10 s (chemiluminescence) or 4–6 s (electrochemical cell) |
| Inspired NO concentration | 40–60 ppm |
| Inspired O2 concentration | Close to 21% |
The 1/θCO equation adopted was empirically derived by Guénard et al. from in vivo single-breath measurements at two alveolar PO2 levels.[27] This equation shows reasonable agreement with several in vitro–derived equations, including those of Roughton and Forster (1957), Forster (1987), and Holland (1969).[2]
Quality control
The ERS Task Force specified the following quality control procedures:[2]
| Item | Requirement |
|---|---|
| Analyser zeroing | Before each test; measure zero level after each test |
| Volume calibration | Daily, with a validated 3-L syringe (accuracy ±2.5%) |
| Leak testing | Weekly or whenever problems are suspected |
| Biological control | Weekly; a healthy nonsmoker whose DL,NO and DL,CO week-to-week should remain within 20 and 5 mL·min−1·mmHg−1, respectively |
| Linearity testing | Monthly, by serial dilution of known gas concentrations |
| Nonlinearity tolerance | ≤1.0% of full scale for NO, CO, and tracer gas analysers |
| Drift limits (over 30 s) | ≤10 ppm CO; ≤1 ppm NO; ≤0.5% tracer gas |
Repeatability and reproducibility
Within a given testing session, two trials whose DL,NO and DL,CO differ by less than 17 and 3.2 mL·min−1·mmHg−1, respectively, are considered acceptable. Week-to-week or month-to-month reproducibility values for DL,NO and DL,CO are 20 and 4.9 mL·min−1·mmHg−1, respectively. A week-to-week change exceeding these thresholds has only a 5% probability of not representing a real change.[2]
Importantly, DL,NO shows a smaller percentage difference between intra-session and inter-session variability (15%) compared to DL,CO (35%), suggesting that DL,NO is a more stable measure over time and that the majority of its variability is within-session rather than between sessions.[2]
No more than eight 5-second breathhold manoeuvres or six 10-second breathhold manoeuvres should be performed in a single session, to avoid excessive rises in carboxyhaemoglobin that could impair DL,CO measurements.[2]
Reference equations
2017 ERS Task Force equations
The original 2017 ERS Task Force reference equations were developed by pooling de-identified data from 490 healthy white adults (248 males, 242 females; age 18–93 years) from three published studies, with a mean breathhold time of approximately 6 seconds.[2][28][29][30] However, approximately 75% of the pooled data in those three studies were collected using the Hyp'Air Compact device (Medisoft, Sorinnes, Belgium), and a 2021 randomised crossover study demonstrated that DL,NO measured by the Hyp'Air was on average 24 mL·min−1·mmHg−1 (17%) higher than by the Jaeger MasterScreen (Vyaire Medical), whereas simultaneously measured DL,CO differed by only 1%.[31] Because the 2017 equations did not account for the testing device, they may introduce systematic bias into predicted DL,NO values depending on the equipment used.
Updated 2022 device-specific equations
In 2022, Zavorsky and Cao published updated reference equations by pooling de-identified data from five studies (530 females, 546 males; age 5–95 years; BMI 12.4–39.0 kg/m2) measured on two commercially available devices: the Jaeger MasterScreen Pro (CareFusion/Vyaire Medical) and the Hyp'Air Compact (Medisoft).[32] The study confirmed that the Hyp'Air Compact measured DL,NO approximately 16–20 mL·min−1·mmHg−1 (13–16%) higher and VA approximately 0.2–0.4 L (6–8%) higher than the Jaeger MasterScreen Pro, whereas DL,CO was similar between devices after controlling for altitude.[32] Segmented (piecewise) linear regression equations with a single age-squared breakpoint were developed and shown to have similar prediction accuracy to generalised additive models of location, scale and shape (GAMLSS), as used by the Global Lung Function Initiative.[32]
The segmented regression equations use age2 as the non-linear covariate, with two line segments connected at a single breakpoint (approximately 22–24 years for DL,CO and DL,NO; approximately 27–30 years for VA). For each variable, there is an equation for the growth phase (below the breakpoint) and a separate equation for the decline phase (above the breakpoint). Height, altitude (for DL,CO), and pulmonary function testing (PFT) device (for DL,NO and VA) are included as covariates where significant.[32]
| Sex | Age range | Intercept | Age2 coefficient | Height (cm) | Altitude (m) | Adj. R2 | RSE |
|---|---|---|---|---|---|---|---|
| Females | 5.0–24.2 years | −11.82 | +0.01534 | 0.183 | 0.0041 | 0.76 | 2.12 |
| 24.3–95.0 years | −1.54 | −0.0018 | 3.13 | ||||
| Males | 5.0–22.6 years | −15.22 | +0.0323 | 0.206 | 0.0041 | 0.80 | 2.62 |
| 22.7–95.0 years | +2.50 | −0.00246 | 4.35 |
| Sex | Age range | Intercept | Age2 coefficient | Height (cm) | PFT equipment[lower-alpha 1] | Adj. R2 | RSE |
|---|---|---|---|---|---|---|---|
| Females | 5.0–22.5 years | −66.43 | +0.0616 | 0.947 | 15.17 | 0.79 | 8.60 |
| 22.6–95.0 years | −30.74 | −0.00832 | 13.63 | ||||
| Males | 5.0–22.1 years | −87.15 | +0.1375 | 1.086 | 18.00 | 0.83 | 11.81 |
| 22.2–95.0 years | −14.02 | −0.012 | 19.25 |
| Sex | Age range | Intercept | Age2 coefficient | Height (cm) | PFT equipment[lower-alpha 2] | Adj. R2 | RSE |
|---|---|---|---|---|---|---|---|
| Females | 5.0–30.2 years | −4.16 | +0.00132 | 0.050 | 0.2545 | 0.80 | 0.39 |
| 30.3–95.0 years | −2.79 | −0.00018 | 0.58 | ||||
| Males | 5.0–26.9 years | −5.64 | +0.00265 | 0.060 | 0.241 | 0.86 | 0.46 |
| 27.0–95.0 years | −3.61 | −0.00013 | 0.73 |
- ↑ For the PFT equipment covariate, 1 = Hyp'Air Compact (Medisoft) and 0 = Jaeger MasterScreen (Vyaire Medical). The Hyp'Air Compact measured DL,NO approximately 15–18 mL·min−1·mmHg−1 higher than the Jaeger MasterScreen Pro.
- ↑ For the PFT equipment covariate, 1 = Hyp'Air Compact and 0 = Jaeger MasterScreen. The Hyp'Air Compact measured VA approximately 0.24–0.25 L higher than the Jaeger MasterScreen Pro.
The lower limit of normal (LLN) is calculated as the predicted value minus 1.645 × RSE (corresponding to the 5th percentile). For example, a 27-year-old man, 180 cm tall, measured on the Hyp'Air Compact would have a predicted VA of −0.00013·(272) + 0.060·(180) + 0.241 − 3.61 = 7.34 L, with an LLN of 7.34 − (0.73 × 1.645) = 6.14 L.[32]
Classification of impairment
Zavorsky and Cao proposed a z-score–based classification system for the severity of diffusion impairment, applicable to DL,NO, DL,CO and VA, since the percentage of predicted value corresponding to the LLN varies with age:[32]
| Category | z-score range | Approximate % predicted (± 3%) |
|---|---|---|
| Severe decrease | ≤ −5.01 | ≤ 41% |
| Moderate decrease | −5.00 to −3.51 | 42–59% |
| Mild decrease | −3.50 to −1.645[lower-alpha 1] | 60–80% |
| Normal | −1.645 to +1.645 | 81–119% |
| Increased | > +1.645 | > 119% |
- ↑ For subjects with suspected or prior evidence of lung disease, the LLN is set at the 5th percentile (z = −1.645). For screening and case-finding purposes, the LLN is set at the 2.5th percentile (z = −1.96), in which case the mild decrease range becomes z = −3.50 to −1.96 (approximately 60–77% predicted).
Applicability and limitations of current reference equations
Both the 2017 ERS and the updated 2022 reference equations were derived from white European and North American subjects. Genetic ancestry is an important covariate in DL,NO prediction. In a pilot study of 59 healthy African American adults matched to white controls for sex, age and height, race accounted for approximately 5–10% of total variance in DL,NO and DL,CO; after controlling for other covariates, Black subjects had a DL,NO that was 12.4 mL·min−1·mmHg−1 lower and a DL,CO that was 3.9 mL·min−1·mmHg−1 lower than white subjects, with much of this difference attributable to a 0.6 L lower VA.[33] Using white-derived reference equations in Black populations risks falsely diagnosing lung disease.
In a larger study of 392 Mexican Hispanics (ages 5–78) compared with 1,056 white subjects, excluding race as a covariate increased the root mean square error by 61% for DL,NO and 18% for DL,CO; race-neutral equations produced false positive rates of 3–6% in Mexican Hispanics and false negative rates of 20–49% in white subjects relative to race-specific equations.[34] Additionally, habitual residence at moderate altitude (2,240 m) increased DL,NO by approximately 7 mL·min−1·mmHg−1 and DL,CO by approximately 4 mL·min−1·mmHg−1 compared with sea-level residents, indicating that altitude should also be incorporated into reference models.[34] Reference equations for other genetic ancestries and diverse geographic populations need to be developed.
Results should be presented as absolute values, percent predicted, and with the corresponding LLN, upper limit of normal (ULN), and z-score.[32] Until between-device discrepancies are resolved by manufacturers, results should be interpreted using device-specific reference values.[32][31]
Limitations and future directions
Approximately 40 years after DL,NO was first measured, five key challenges have been identified that must be addressed to move the test from a niche research tool to routine clinical practice.[21]
Challenge 1 – A thin commercial ecosystem and regulatory inertia. Devices capable of simultaneous NO–CO measurements exist, but manufacturers are few and platforms differ in sensor technology (chemiluminescence, electrochemical, laser), response time, cross-sensitivities, calibration routines, gas mixtures and calculation algorithms. These differences shift absolute values and complicate portability of reference equations across devices. In the United States, the absence of Food and Drug Administration clearance for NO–CO systems and limited clinician awareness have further restricted DL,NO to research use. Multi-site method-comparison trials that include device brand as a covariate, shared quality-assurance protocols, and regulatory-ready analytical validation packages are needed. Laboratories should explicitly report device brand, analyser type, software version and breathhold time in manuscripts and clinical reports.[21]
Challenge 2 – Reference equations must account for altitude and genetic ancestry. Diffusing capacity varies across populations and environments. In Mexican Hispanics, habitual residence at approximately 2,240 m increased DL,NO, DL,CO and VA compared with lowlanders, and models that included altitude and population terms reduced error and misclassification relative to race-neutral models.[34] Complementary work argued specifically for race-specific DL,NO equations to reduce false positives and false negatives in clinical practice.[33] Where device- and timing-specific equations for particular populations already exist, laboratories should report z-scores that incorporate altitude and ancestry when they measurably improve calibration, and consider parallel reporting of race-neutral values when available.[21]
Challenge 3 – Building global DL,NO reference equations. The field needs multi-centre, multi-ancestry DL,NO–DL,CO datasets that explicitly encode device brand, altitude and breathhold time. A standing consortium could harmonise protocols, pool raw data and generate globally applicable equations with clear device and timing annotations. Current validated datasets exist for Mexican Hispanics,[34] Black adults[33] and white populations,[32] but larger samples are required in Black populations and broader ancestry representation is needed to minimise misclassification and facilitate international adoption. A minimal data standard should include triplicate DL,NO–DL,CO manoeuvres with 5–6 s breathhold time, accepted quality criteria, haemoglobin values, body size, age and sex, and reporting of spirometry and lung volumes.[21]
Challenge 4 – Two gases are better than one. Adding DL,NO to DL,CO materially improves case detection. An individual-participant data meta-analysis in heart failure found that the double-diffusion approach enhanced classification compared with DL,CO alone. In post-COVID cohorts, the summed DL,NO + DL,CO z-score outperformed DL,CO alone for identifying prior disease, and in smokers with emphysema, hierarchical partitioning showed that DL,NO z-scores contributed uniquely to model fit beyond spirometry and lung volumes, whereas DL,CO z-scores did not add further useful information. Whenever a DL,CO test is clinically indicated and equipment is available, DL,NO should be measured in the same manoeuvre and both values reported together with their ratio.[21]
Challenge 5 – Reconciling component-splitting with time-based kinetics. Clinicians now encounter two frameworks for interpreting diffusing capacity. The classic Roughton–Forster framework treats total resistance as two serial terms—membrane (1/DM) and blood (1/(θ·VC))—and has been used for back-calculations of DM and VC from paired DL,NO–DL,CO. A time-based diffusion–reaction framework instead derives uptake from first principles as the sum of an equivalent diffusion time and a reaction time. Under this model, DL,NO behaves as surface absorption dominated by the membrane–plasma path and accessible red cell surface, whereas DL,CO behaves as volume absorption that scales with haematocrit and red cell volume. In clinical reports, it is recommended to present DL,NO, DL,CO, VA and the DL,NO/DL,CO ratio as observables and to flag any DM–VC values as model-dependent.[21]
See also
- Diffusing capacity
- Pulmonary function testing
- Nitric oxide
- Carbon monoxide
- Roughton–Forster equation
- Exhaled nitric oxide
References
- ↑ 1.00 1.01 1.02 1.03 1.04 1.05 1.06 1.07 1.08 1.09 1.10 1.11 1.12 Borland, Colin D.R.; Guénard, Hervé (2017). "The history of the pulmonary diffusing capacity for nitric oxide DL,NO". Respiratory Physiology & Neurobiology 241: 3–6. doi:10.1016/j.resp.2016.11.014.
- ↑ 2.00 2.01 2.02 2.03 2.04 2.05 2.06 2.07 2.08 2.09 2.10 2.11 2.12 2.13 2.14 2.15 2.16 2.17 2.18 2.19 2.20 2.21 2.22 2.23 2.24 2.25 2.26 2.27 2.28 2.29 2.30 2.31 2.32 2.33 2.34 2.35 2.36 2.37 Zavorsky, Gerald S.; Hsia, Connie C.W.; Hughes, J. Michael B.; Borland, Colin D.R.; Guénard, Hervé; van der Lee, Ivo; Steenbruggen, Irene; Naeije, Robert et al. (2017). "Standardisation and application of the single-breath determination of nitric oxide uptake in the lung". European Respiratory Journal 49 (1): 1600962. doi:10.1183/13993003.00962-2016. PMID 28049169.
- ↑ 3.0 3.1 Guénard, H.; Varene, N.; Vaida, P. (1987). "Determination of lung capillary blood volume and membrane diffusing capacity in man by the measurements of NO and CO transfer". Respiratory Physiology 70 (1): 113–120. doi:10.1016/S0034-5687(87)80036-1. PMID 3659606.
- ↑ Clyne, M.A.A.; Thrush, B.A.; Wayne, R.P. (1964). "Kinetics of the chemiluminescent reaction between nitric oxide and ozone". Transactions of the Faraday Society 60: 359–361. doi:10.1039/tf9646000359.
- ↑ Higenbottam, T.; Clark, T.J.; Shipley, M.J.; Rose, G. (1980). "Lung function and symptoms of cigarette smokers related to tar yield and number of cigarettes smoked". Lancet 23: 409–411. doi:10.1016/S0140-6736(80)90973-X. PMID 6105497.
- ↑ Borland, C.; Chamberlain, A.; Higenbottam, T.W. (1983). "The fate of inhaled nitric oxide". Clinical Science 65: 37P.
- ↑ Borland, C.; Chamberlain, A.; Higenbottam, T. (1984). "Is the measurement of DL,NO a true measurement of membrane diffusing capacity?". Clinical Science 67: 41P.
- ↑ Borland, Colin D.R.; Higenbottam, Tim W. (1989). "A simultaneous single breath measurement of pulmonary diffusing capacity with nitric oxide and carbon monoxide". European Respiratory Journal 2 (1): 56–63. PMID 2707403.
- ↑ 9.0 9.1 Carlsen, E.; Comroe, J.H. Jr (1958). "The rate of uptake of carbon monoxide and of nitric oxide by normal human erythrocytes and experimentally produced spherocytes". Journal of General Physiology 42 (1): 83–107. doi:10.1085/jgp.42.1.83. PMID 13575776.
- ↑ Ignarro, L.J.; Buga, G.M.; Wood, K.S.; Byrns, R.E.; Chaudhuri, G. (1987). "Endothelium-derived relaxing factor produced and released from artery and vein is nitric oxide". Proceedings of the National Academy of Sciences USA 84: 9265–9269. doi:10.1073/pnas.84.24.9265. PMID 2827174.
- ↑ Palmer, R.M.J.; Ferrige, A.G.; Moncada, S. (1987). "Nitric oxide release accounts for the biological activity of endothelium-derived relaxing factor". Nature 327 (6122): 524–526. doi:10.1038/327524a0. PMID 3495737.
- ↑ Forster, R.E. (1987). "Diffusion of gases across the alveolar membrane". in Fishman, A.P.; Farhi, L.E.; Tenney, S.M.. Handbook of Physiology, Section 3: The Respiratory System, Vol IV: Gas Exchange. Washington, DC: American Physiological Society. pp. 71–88.
- ↑ Borland, Colin D.R.; Dunningham, H.; Bottrill, F.; Vuylsteke, A. (2006). "Can a membrane oxygenator be a model for NO and CO transfer?". Journal of Applied Physiology 100 (5): 1527–1538. doi:10.1152/japplphysiol.00977.2005. PMID 16424068.
- ↑ 14.0 14.1 Borland, Colin D.R.; Dunningham, H.; Bottrill, F.; Vuylsteke, A.; Yilmaz, C.; Dane, D.M.; Hsia, Connie C.W. (2010). "Significant blood resistance to nitric oxide transfer in the lung". Journal of Applied Physiology 108 (4): 1052–1060. doi:10.1152/japplphysiol.00998.2009. PMID 20150428.
- ↑ 15.0 15.1 Borland, C.; Bottrill, F.; Jones, A.; Sherwood, C.; Vuylsteke, A. (2014). "The significant blood resistance to lung nitric oxide transfer lies within the red cell". Journal of Applied Physiology 116 (1): 32–41. doi:10.1152/japplphysiol.00959.2013. PMID 24177693.
- ↑ Roughton, F.J.W.; Forster, R.E. (1957). "Relative importance of diffusion and chemical reaction rates in determining rate of exchange of gases in the human lung, with special reference to true diffusing capacity of pulmonary membrane and volume of blood in the lung capillaries". Journal of Applied Physiology 11 (2): 290–302. doi:10.1152/jappl.1957.11.2.290. PMID 13475180.
- ↑ Gibson, Q.H.; Roughton, F.J.W. (1957). "The kinetics and equilibria of the reactions of nitric oxide with sheep haemoglobin". Journal of Physiology 136: 507–524. doi:10.1113/jphysiol.1957.sp005777. PMID 13429516.
- ↑ Wilhelm, E.; Battino, R.; Wilcock, R.J. (1977). "Low-pressure solubility of gases in liquid water". Chemical Reviews 77 (2): 219–262. doi:10.1021/cr60306a003.
- ↑ 19.0 19.1 19.2 19.3 19.4 Kang, Min-Yeong; Grebenkov, Denis; Guénard, Hervé; Katz, Ira; Sapoval, Bernard (2017). "The Roughton–Forster equation for DL,CO and DL,NO re-examined". Respiratory Physiology & Neurobiology 241: 62–71. doi:10.1016/j.resp.2016.12.014. PMID 28049017.
- ↑ 20.0 20.1 20.2 Sapoval, Bernard; Kang, Min-Yeong; Dinh-Xuan, Anh Tuan (2021). "Modeling of gas exchange in the lungs". Comprehensive Physiology 11 (1): 1289–1314. doi:10.1002/cphy.c190019. PMID 33295657.
- ↑ 21.0 21.1 21.2 21.3 21.4 21.5 21.6 21.7 Zavorsky, Gerald S. (2025). "Challenges of DLNO 40 years after its invention". Archivos de Bronconeumología. doi:10.1016/j.arbres.2025.11.007.
- ↑ Zavorsky, G.S.; Lands, L.C. (2005). "Lung diffusion capacity for nitric oxide and carbon monoxide is impaired similarly following short-term graded exercise". Nitric Oxide 12 (1): 31–38. doi:10.1016/j.niox.2004.11.002. PMID 15631944.
- ↑ 23.0 23.1 Zavorsky, Gerald S.; Agostoni, Piergiuseppe (2024). "Two is better than one: the double diffusion technique in classifying heart failure". ERJ Open Research 10 (1): 00644-2023. doi:10.1183/23120541.00644-2023. PMID 38226067.
- ↑ 24.0 24.1 Zavorsky, Gerald S.; Barisione, Giovanni; Gille, Thomas; Villasante, Carlos; Seccombe, Leigh M.; Farah, Claude S.; Núñez-Fernández, Marina; Shan, Shi Huh Samuel et al. (2025). "Enhanced detection of patients with previous COVID-19: Superiority of the double diffusion technique". BMJ Open Respiratory Research 12 (1). doi:10.1136/bmjresp-2024-002561.
- ↑ 25.0 25.1 Zavorsky, Gerald S.; Dal Negro, Roberto W.; van der Lee, Ivo; Preisser, Alexandra M. (2025). "Emphysema detection in smokers: diffusing capacity for nitric oxide beats diffusing capacity for carbon monoxide-based models". Chronic Obstructive Pulmonary Diseases: Journal of the COPD Foundation 12 (6): 500–511. doi:10.15326/jcopdf.2025.0645.
- ↑ Jones, R.S.; Meade, F. (1961). "A theoretical and experimental analysis of anomalies in the estimation of pulmonary diffusing capacity by the single breath method". Quarterly Journal of Experimental Physiology 46: 131–143. doi:10.1113/expphysiol.1961.sp001525. PMID 13726417.
- ↑ Guénard, H.J.; Martinot, J.B.; Martin, S.; Maury, B.; Lalande, S.; Kays, C. (2016). "In vivo estimates of NO and CO conductance for haemoglobin and for lung transfer in humans". Respiratory Physiology & Neurobiology 228: 1–8. doi:10.1016/j.resp.2016.03.005. PMID 26993367.
- ↑ van der Lee, I.; Zanen, P.; Stigter, N.; van den Bosch, J.M.M.; Lammers, J.J.W. (2007). "Diffusing capacity for nitric oxide: reference values and dependence on alveolar volume". Respiratory Medicine 101 (7): 1579–1584. doi:10.1016/j.rmed.2006.12.009. PMID 17254764.
- ↑ Aguilaniu, B.; Maitre, J.; Glénet, S.; Gegout-Petit, A.; Guénard, H. (2008). "European reference equations for CO and NO lung transfer". European Respiratory Journal 31 (5): 1091–1097. doi:10.1183/09031936.00063207. PMID 18216064.
- ↑ Zavorsky, G.S.; Cao, J.; Murias, J.M. (2008). "Reference values of pulmonary diffusing capacity for nitric oxide in an adult population". Nitric Oxide 18 (1): 70–79. doi:10.1016/j.niox.2007.10.002. PMID 18036856.
- ↑ 31.0 31.1 Radtke, Thomas; de Groot, Quintin; Haile, Sarah R.; Maggi, Marion; Hsia, Connie C.W.; Dressel, Holger (2021). "Lung diffusing capacity for nitric oxide measured by two commercial devices: a randomised crossover comparison in healthy adults". ERJ Open Research 7 (3): 00193-2021. doi:10.1183/23120541.00193-2021.
- ↑ 32.00 32.01 32.02 32.03 32.04 32.05 32.06 32.07 32.08 32.09 32.10 32.11 32.12 Zavorsky, Gerald S.; Cao, Jiguo (2022). "Reference equations for pulmonary diffusing capacity using segmented regression show similar predictive accuracy as GAMLSS models". BMJ Open Respiratory Research 9 (1). doi:10.1136/bmjresp-2021-001087. PMID 35172984.
- ↑ 33.0 33.1 33.2 Zavorsky, Gerald S.; Almamary, Ahmad S.; Alqahtani, Mobarak K.; Shan, Shi Huh Samuel; Gardenhire, Douglas S. (2021). "The need for race-specific reference equations for pulmonary diffusing capacity for nitric oxide". BMC Pulmonary Medicine 21 (1): 232. doi:10.1186/s12890-021-01591-7. PMID 34256739.
- ↑ 34.0 34.1 34.2 34.3 Gochicoa-Rangel, Laura; De-Los-Santos-Martínez, Ada; Reyes-García, Alejandro; Martínez-Briseño, David; Vargas, Mario H.; Lechuga-Trejo, Irma; Guzmán-Valderrábano, Carlos; Torre-Bouscoulet, Luis et al. (2024). "Reference equations for DLNO and DLCO in Mexican Hispanics: influence of altitude and race". BMJ Open Respiratory Research 11 (1). doi:10.1136/bmjresp-2024-002341. PMID 39401975.
Further reading
- Hughes, J.M.B.; van der Lee, I. (2013). "The TL,NO/TL,CO ratio in pulmonary function test interpretation". European Respiratory Journal 41 (2): 453–461. doi:10.1183/09031936.00082212. PMID 22936707.
- Martinot, J.B. et al. (2016). "Nitrogen monoxide and carbon monoxide transfer interpretation: state of the art". Clinical Physiology and Functional Imaging. doi:10.1111/cpf.12316.
External links
- European Respiratory Society — ERS official website
- Full text of 2017 ERS Task Force document — European Respiratory Journal
- Full text of 2022 updated reference equations — BMJ Open Respiratory Research
- Race-specific reference equations for DLNO — BMC Pulmonary Medicine
- Reference equations for Mexican Hispanics: altitude and race — BMJ Open Respiratory Research
- Challenges of DLNO 40 years after its invention — Archivos de Bronconeumología
