Engineering:Electron microscope
This article needs additional citations for verification. (September 2023) (Learn how and when to remove this template message) |
An electron microscope is a microscope that uses a beam of electrons as a source of illumination. They use electron optics that are analogous to the glass lenses of an optical light microscope to control the electron beam, for instance focusing them to produce magnified images or electron diffraction patterns. As the wavelength of an electron can be up to 100,000 times smaller than that of visible light, electron microscopes have a much higher resolution of about 0.1 nm, which compares to about 200 nm for light microscopes. Electron microscope may refer to:
- Transmission electron microscopy (TEM) where swift electrons go through a thin sample
- Scanning transmission electron microscopy (STEM) which is similar to TEM with a scanned electron probe
- Scanning electron microscope (SEM) which is similar to STEM, but with thick samples
- Electron microprobe similar to a SEM, but more for chemical analysis
- Ultrafast scanning electron microscopy, version of a SEM that can operate very fast
- Low-energy electron microscopy (LEEM), used to image surfaces
- Photoemission electron microscopy (PEEM) which is similar to LEEM using electrons emitted from surfaces by photons
Additional details can be found in the above links. This article contains some general information mainly about transmission electron microscopes.
History
Many developments laid the groundwork of the electron optics used in microscopes.[1] One significant step was the work of Hertz in 1883[2] who made a cathode-ray tube with electrostatic and magnetic deflection, demonstrating manipulation of the direction of an electron beam. Others were focusing of the electrons by an axial magnetic field by Emil Wiechert in 1899,[3] improved oxide-coated cathodes which produced more electrons by Arthur Wehnelt in 1905[4] and the development of the electromagnetic lens in 1926 by Hans Busch.[5] According to Dennis Gabor, the physicist Leó Szilárd tried in 1928 to convince him to build an electron microscope, for which Szilárd had filed a patent.[6]
To this day the issue of who invented the transmission electron microscope is controversial.[7][8][9][10] In 1928, at the Technical University of Berlin, Adolf Matthias (Professor of High Voltage Technology and Electrical Installations) appointed Max Knoll to lead a team of researchers to advance research on electron beams and cathode-ray oscilloscopes. The team consisted of several PhD students including Ernst Ruska. In 1931, Max Knoll and Ernst Ruska[11][12] successfully generated magnified images of mesh grids placed over an anode aperture. The device, a replicate of which is shown in the figure, used two magnetic lenses to achieve higher magnifications, the first electron microscope. (Max Knoll died in 1969, so did not receive a share of the 1986 Nobel prize for the invention of electron microscopes.)
Apparently independent of this effort was work at Siemens-Schuckert by Reinhold Rüdenberg. According to patent law (U.S. Patent No. 2058914[13] and 2070318,[14] both filed in 1932), he is the inventor of the electron microscope, but it is not clear when he had a working instrument. He stated in a very brief article in 1932[15] that Siemens had been working on this for some years before the patents were filed in 1932, claiming that his effort was parallel to the university development. He died in 1961, so similar to Max Knoll, was not eligible for a share of the 1986 Nobel prize.
In the following year, 1933, Ruska and Knoll built the first electron microscope that exceeded the resolution of an optical (light) microscope.[16] Four years later, in 1937, Siemens financed the work of Ernst Ruska and Bodo von Borries, and employed Helmut Ruska, Ernst's brother, to develop applications for the microscope, especially with biological specimens.[16][17] Also in 1937, Manfred von Ardenne pioneered the scanning electron microscope.[18] Siemens produced the first commercial electron microscope in 1938.[19] The first North American electron microscopes were constructed in the 1930s, at the Washington State University by Anderson and Fitzsimmons [20] and at the University of Toronto by Eli Franklin Burton and students Cecil Hall, James Hillier, and Albert Prebus. Siemens produced a transmission electron microscope (TEM) in 1939.[21] Although current transmission electron microscopes are capable of two million times magnification, as scientific instruments they remain similar but with improved optics.
Wavelength
File:Transmission Electron Microscope operating principle.ogg In a typical electron gun, individual electrons, which have an elementary charge [math]\displaystyle{ e }[/math] (about [math]\displaystyle{ -1.6\times10^{-19} }[/math] coulombs) and a mass [math]\displaystyle{ m }[/math] (about [math]\displaystyle{ 9.1\times 10^{-31} }[/math] kg), with a potential of [math]\displaystyle{ V }[/math] volts, have an energy amount of [math]\displaystyle{ e\cdot V }[/math] joules. The wavelength is[22]
- [math]\displaystyle{ \lambda=\frac{hc}{\sqrt{eV(2mc^2+eV)}} }[/math],
where [math]\displaystyle{ c }[/math] is the speed of light in vacuum (about [math]\displaystyle{ c=3\times 10^8 }[/math] m/s). See electron diffraction for a full explanation.
Types
Transmission electron microscope (TEM)
The original form of the electron microscope, the transmission electron microscope (TEM), uses a high voltage electron beam to illuminate the specimen and create an image. An electron beam is produced by an electron gun, with the electrons typically at 40 to 400 keV, focused by electromagnetic lenses, and transmitted through the specimen. When it emerges from the specimen, the electron beam carries information about the structure of the specimen that is magnified by lenses of the microscope. The spatial variation in this information (the "image") may be viewed by projecting the magnified electron image onto a detector. For example, the image may be viewed directly by an operator using a fluorescent viewing screen coated with a phosphor or scintillator material such as zinc sulfide. A high-resolution phosphor may also be coupled by means of a lens optical system or a fibre optic light-guide to the sensor of a digital camera. Direct electron detectors have no scintillator and are directly exposed to the electron beam, which addresses some of the limitations of scintillator-coupled cameras.[23]
The resolution of TEMs is limited primarily by spherical aberration, but a new generation of hardware correctors can reduce spherical aberration to increase the resolution in high-resolution transmission electron microscopy (HRTEM) to below 0.5 angstrom (50 picometres),[24] enabling magnifications above 50 million times.[25] The ability of HRTEM to determine the positions of atoms within materials is useful for nano-technologies research and development.[26]
Transmission electron microscopes are often used in electron diffraction mode. The advantages of electron diffraction over X-ray crystallography are that the specimen need not be a single crystal or even a polycrystalline powder.[citation needed]
Scanning transmission electron microscope (STEM)
The STEM rasters a focused incident probe across a specimen. The high resolution of the TEM is thus possible in STEM. The focusing action (and aberrations) occur before the electrons hit the specimen in the STEM, but afterward in the TEM. The STEMs use of SEM-like beam rastering simplifies annular dark-field imaging, and other analytical techniques, but also means that image data is acquired in serial rather than in parallel fashion.[citation needed]
Scanning electron microscope (SEM)
File:Scanning Electron Microscope.ogv
The SEM produces images by probing the specimen with a focused electron beam that is scanned across the specimen (raster scanning). When the electron beam interacts with the specimen, it loses energy by a variety of mechanisms. The lost energy is converted into alternative forms such as heat, emission of low-energy secondary electrons and high-energy backscattered electrons, light emission (cathodoluminescence) or X-ray emission, all of which provide signals carrying information about the properties of the specimen surface, such as its topography and composition.[citation needed] The image displayed by an SEM maps the varying intensity of any of these signals into the image in a position corresponding to the position of the beam on the specimen when the signal was generated. In the SEM image of an ant shown, the image was constructed from signals produced by a secondary electron detector, the normal or conventional imaging mode in most SEMs.[citation needed]
Generally, the image resolution of an SEM is lower than that of a TEM. However, because the SEM images the surface of a sample rather than its interior, the electrons do not have to travel through the sample. This reduces the need for extensive sample preparation to thin the specimen to electron transparency. The SEM also has a great depth of field, and so can produce images that are good representations of the three-dimensional surface shape of the sample.[citation needed]
In their most common configurations, electron microscopes produce images with a single brightness value per pixel, with the results usually rendered in greyscale.[27] However, often these images are then colourized through the use of feature-detection software, or simply by hand-editing using a graphics editor. This may be done to clarify structure or for aesthetic effect and generally does not add new information about the specimen.[28]
Sample preparation for TEM
Materials to be viewed in a transmission electron microscope may require processing to produce a suitable sample. The technique required varies depending on the specimen and the analysis required:
- Chemical fixation – for biological specimens this aims to stabilize the specimen's mobile macromolecular structure by chemical crosslinking of proteins with aldehydes such as formaldehyde and glutaraldehyde, and lipids with osmium tetroxide.[29]
- Cryofixation – freezing a specimen so that the water forms vitreous (non-crystalline) ice. This preserves the specimen in a snapshot of its native state. Methods to achieve this vitrification include plunge freezing rapidly in liquid ethane, and high pressure freezing. An entire field called cryo-electron microscopy has branched from this technique. With the development of cryo-electron microscopy of vitreous sections (CEMOVIS)[30] and cryo-focused ion beam milling of lamellae,[31] it is now possible to observe samples from virtually any biological specimen close to its native state.
- Dehydration – replacement of water with organic solvents such as ethanol or acetone, followed by critical point drying or infiltration with embedding resins. See also freeze drying.[citation needed]
- Embedding, biological specimens – after dehydration, tissue for observation in the transmission electron microscope is embedded so it can be sectioned ready for viewing. To do this the tissue is passed through a 'transition solvent' such as propylene oxide (epoxypropane) or acetone and then infiltrated with an epoxy resin such as Araldite, Epon, or Durcupan;[32] tissues may also be embedded directly in water-miscible acrylic resin. After the resin has been polymerized (hardened) the sample is sectioned by ultramicrotomy and stained.[citation needed]
- Embedding, materials – after embedding in resin, the specimen is usually ground and polished to a mirror-like finish using ultra-fine abrasives.[citation needed]
- Freeze-fracture replica immunogold labeling (FRIL) – the freeze-fracture method has been modified to allow the identification of the components of the fracture face by immunogold labeling. Instead of removing all the underlying tissue of the thawed replica as the final step before viewing in the microscope the tissue thickness is minimized during or after the fracture process. The thin layer of tissue remains bound to the metal replica so it can be immunogold labeled with antibodies to the structures of choice. The thin layer of the original specimen on the replica with gold attached allows the identification of structures in the fracture plane.[33] There are also related methods which label the surface of etched cells[34] and other replica labeling variations.[35]
- Ion beam milling – thins samples until they are transparent to electrons by firing ions (typically argon) at the surface from an angle and sputtering material from the surface. A subclass of this is focused ion beam milling, where gallium ions are used to produce an electron transparent membrane or 'lamella' in a specific region of the sample, for example through a device within a microprocessor or a focused ion beam SEM. Ion beam milling may also be used for cross-section polishing prior to analysis of materials that are difficult to prepare using mechanical polishing.[citation needed]
- Negative stain – suspensions containing nanoparticles or fine biological material (such as viruses and bacteria) are briefly mixed with a dilute solution of an electron-opaque solution such as ammonium molybdate, uranyl acetate (or formate), or phosphotungstic acid.[citation needed] This mixture is applied to an EM grid, pre-coated with a plastic film such as formvar, blotted, then allowed to dry. Viewing of this preparation in the TEM should be carried out without delay for best results. The method is important in microbiology for fast but crude morphological identification, but can also be used as the basis for high-resolution 3D reconstruction using EM tomography methodology when carbon films are used for support. Negative staining is also used for observation of nanoparticles.[citation needed]
- Sectioning – produces thin slices of the specimen, semitransparent to electrons. These can be cut using ultramicrotomy on an ultramicrotome with a glass or diamond knife to produce ultra-thin sections about 60–90 nm thick. Disposable glass knives are also used because they can be made in the lab and are much cheaper. Sections can also be created in situ by milling in a focused ion beam SEM, where the section is known as a lamella.[31]
- Staining – uses heavy metals such as lead, uranium or tungsten to scatter imaging electrons and thus give contrast between different structures, since many (especially biological) materials are nearly "transparent" to electrons (weak phase objects). In biology, specimens can be stained "en bloc" before embedding and also later after sectioning. Typically thin sections are stained for several minutes with an aqueous or alcoholic solution of uranyl acetate followed by aqueous lead citrate.[36]
EM workflows
Early electron microscopy of biological specimens was often descriptive, making use of the newly available higher resolution.[37] This is still the case for various applications, such as diagnostic electron microscopy.
However, electron microscopes are now frequently used in more complex workflows, with each workflow typically using multiple technologies to enable more complex and/or more quantitative analyses of a sample. A few examples are outlined below, but this should not be considered an exhaustive list. The choice of workflow will be highly dependent on the application and the requirements of the corresponding scientific questions, such as resolution, volume, nature of the target molecule, etc.
For example, images from light and electron microscopy of the same region of a sample can be overlaid to correlate the data from the two modalities. This is commonly used to provide higher resolution contextual EM information about a fluorescently labelled structure. This correlative light and electron microscopy (CLEM)[38] is one of a range of correlative workflows now available. Another example is high resolution mass spectrometry (ion microscopy), which has been used to provide correlative information about subcellular antibiotic localisation,[39] data that would be difficult to obtain by other means.[citation needed]
The initial role of electron microscopes in imaging two-dimensional slices (TEM) or a specimen surface (SEM with secondary electrons) has also increasingly expanded into the depth of samples.[40] An early example of these ‘volume EM’ workflows was simply to stack TEM images of serial sections cut through a sample. The next development was virtual reconstruction of a thick section (200-500 nm) volume by backprojection of a set of images taken at different tilt angles - TEM tomography.[41]
Serial imaging for volume EM
To acquire volume EM datasets of larger depths than TEM tomography (micrometers or millimeters in the z axis), a series of images taken through the sample depth can be used. For example, ribbons of serial sections can be imaged in a TEM as described above, and when thicker sections are used, serial TEM tomography can be used to increase the z-resolution. More recently, back scattered electron (BSE) images can be acquired of a larger series of sections collected on silicon wafers, known as SEM array tomography.[42][43] An alternative approach is to use BSE SEM to image the block surface instead of the section, after each section has been removed. By this method, an ultramicrotome installed in an SEM chamber can increase automation of the workflow; the specimen block is loaded in the chamber and the system programmed to continuously cut and image through the sample. This is known as serial block face SEM.[44] A related method uses focused ion beam milling instead of an ultramicrotome to remove sections. In these serial imaging methods, the output is essentially a sequence of images through a specimen block that can be digitally aligned in sequence and thus reconstructed into a volume EM dataset. The increased volume available in these methods has expanded the capability of electron microscopy to address new questions,[40] such as mapping neural connectivity in the brain,[45] and membrane contact sites between organelles.[46]
Disadvantages
Electron microscopes are expensive to build and maintain. Microscopes designed to achieve high resolutions must be housed in stable buildings (sometimes underground) with special services such as magnetic field canceling systems.[citation needed]
The samples largely have to be viewed in vacuum, as the molecules that make up air would scatter the electrons. An exception is liquid-phase electron microscopy[47] using either a closed liquid cell or an environmental chamber, for example, in the environmental scanning electron microscope, which allows hydrated samples to be viewed in a low-pressure (up to 20 Torr or 2.7 kPa) wet environment. Various techniques for in situ electron microscopy of gaseous samples have been developed.[48]
Scanning electron microscopes operating in conventional high-vacuum mode usually image conductive specimens; therefore non-conductive materials require conductive coating (gold/palladium alloy, carbon, osmium, etc.). The low-voltage mode of modern microscopes makes possible the observation of non-conductive specimens without coating. Non-conductive materials can be imaged also by a variable pressure (or environmental) scanning electron microscope.[citation needed]
Small, stable specimens such as carbon nanotubes, diatom frustules and small mineral crystals (asbestos fibres, for example) require no special treatment before being examined in the electron microscope. Samples of hydrated materials, including almost all biological specimens, have to be prepared in various ways to stabilize them, reduce their thickness (ultrathin sectioning) and increase their electron optical contrast (staining). These processes may result in artifacts, but these can usually be identified by comparing the results obtained by using radically different specimen preparation methods. Since the 1980s, analysis of cryofixed, vitrified specimens has also become increasingly used by scientists, further confirming the validity of this technique.[49][50][51]
See also
- List of materials analysis methods
- Electron diffraction
- Electron energy loss spectroscopy (EELS)
- Energy filtered transmission electron microscopy (EFTEM)
- Environmental scanning electron microscope (ESEM)
- Immune electron microscopy
- In situ electron microscopy
- Low-energy electron microscopy
- Microscope image processing
- Microscopy
- Nanotechnology
- Scanning confocal electron microscopy
- Scanning electron microscope (SEM)
- Thin section
- Transmission Electron Aberration-Corrected Microscope
References
- ↑ Calbick, C. J. (1944). "Historical Background of Electron Optics" (in en). Journal of Applied Physics 15 (10): 685–690. doi:10.1063/1.1707371. ISSN 0021-8979. Bibcode: 1944JAP....15..685C. http://aip.scitation.org/doi/10.1063/1.1707371.
- ↑ Hertz, Heinrich (2019), "Introduction to Heinrich Hertz's Miscellaneous Papers (1895) by Philipp Lenard", Heinrich Rudolf Hertz (1857-1894) (Routledge): pp. 87–88, doi:10.4324/9780429198960-4, ISBN 978-0-429-19896-0, http://dx.doi.org/10.4324/9780429198960-4, retrieved 2023-02-24
- ↑ Wiechert, E. (1899). "Experimentelle Untersuchungen über die Geschwindigkeit und die magnetische Ablenkbarkeit der Kathodenstrahlen" (in de). Annalen der Physik und Chemie 305 (12): 739–766. doi:10.1002/andp.18993051203. Bibcode: 1899AnP...305..739W. https://onlinelibrary.wiley.com/doi/10.1002/andp.18993051203.
- ↑ Wehnelt, A. (1905). "X. On the discharge of negative ions by glowing metallic oxides, and allied phenomena" (in en). The London, Edinburgh, and Dublin Philosophical Magazine and Journal of Science 10 (55): 80–90. doi:10.1080/14786440509463347. ISSN 1941-5982. https://www.tandfonline.com/doi/full/10.1080/14786440509463347.
- ↑ Busch, H. (1926). "Berechnung der Bahn von Kathodenstrahlen im axialsymmetrischen elektromagnetischen Felde" (in de). Annalen der Physik 386 (25): 974–993. doi:10.1002/andp.19263862507. Bibcode: 1926AnP...386..974B. https://onlinelibrary.wiley.com/doi/10.1002/andp.19263862507.
- ↑ Dannen, Gene (1998) Leo Szilard the Inventor: A Slideshow (1998, Budapest, conference talk). dannen.com
- ↑ Mulvey, T (1962). "Origins and historical development of the electron microscope". British Journal of Applied Physics 13 (5): 197–207. doi:10.1088/0508-3443/13/5/303. ISSN 0508-3443. https://iopscience.iop.org/article/10.1088/0508-3443/13/5/303.
- ↑ Tao, Yaping (2018). "A Historical Investigation of the Debates on the Invention and Invention Rights of Electron Microscope" (in en). Proceedings of the 3rd International Conference on Contemporary Education, Social Sciences and Humanities (ICCESSH 2018). Atlantis Press. 1438–1441. doi:10.2991/iccessh-18.2018.313. ISBN 978-94-6252-528-3. https://www.atlantis-press.com/proceedings/iccessh-18/25898208.
- ↑ Freundlich, Martin M. (1963). "Origin of the Electron Microscope: The history of a great invention, and of a misconception concerning the inventors, is reviewed." (in en). Science 142 (3589): 185–188. doi:10.1126/science.142.3589.185. ISSN 0036-8075. PMID 14057363. https://www.science.org/doi/10.1126/science.142.3589.185.
- ↑ Rüdenberg, Reinhold (2010), Origin and Background of the Invention of the Electron Microscope, Advances in Imaging and Electron Physics, 160, Elsevier, pp. 171–205, doi:10.1016/s1076-5670(10)60005-5, ISBN 9780123810175, http://dx.doi.org/10.1016/s1076-5670(10)60005-5, retrieved 2023-02-11.
- ↑ Knoll, M.; Ruska, E. (1932). "Beitrag zur geometrischen Elektronenoptik. I". Annalen der Physik 404 (5): 607–640. doi:10.1002/andp.19324040506. ISSN 0003-3804. Bibcode: 1932AnP...404..607K. http://dx.doi.org/10.1002/andp.19324040506.
- ↑ Knoll, M.; Ruska, E. (1932). "Das Elektronenmikroskop" (in de). Zeitschrift für Physik 78 (5–6): 318–339. doi:10.1007/BF01342199. ISSN 1434-6001. Bibcode: 1932ZPhy...78..318K. http://link.springer.com/10.1007/BF01342199.
- ↑ Rüdenberg, Reinhold. "Apparatus for producing images of objects". https://image-ppubs.uspto.gov/dirsearch-public/print/downloadPdf/2058914.
- ↑ Rüdenberg, Reinhold. "Apparatus for producing images of objects". https://image-ppubs.uspto.gov/dirsearch-public/print/downloadPdf/2070318.
- ↑ Rodenberg, R. (1932). "Elektronenmikroskop" (in de). Die Naturwissenschaften 20 (28): 522. doi:10.1007/BF01505383. ISSN 0028-1042. Bibcode: 1932NW.....20..522R. http://link.springer.com/10.1007/BF01505383.
- ↑ 16.0 16.1 Ruska, Ernst (1986). "Ernst Ruska Autobiography". Nobel Foundation. http://nobelprize.org/nobel_prizes/physics/laureates/1986/ruska-autobio.html.
- ↑ Kruger, DH; Schneck, P; Gelderblom, HR (May 2000). "Helmut Ruska and the visualisation of viruses". The Lancet 355 (9216): 1713–1717. doi:10.1016/S0140-6736(00)02250-9. PMID 10905259.
- ↑ Ardenne, M. Von; Beischer, D. (1940). "Untersuchung von Metalloxyd-Rauchen mit dem Universal-Elektronenmikroskop" (in de). Zeitschrift für Elektrochemie und Angewandte Physikalische Chemie 46 (4): 270–277. doi:10.1002/bbpc.19400460406. https://onlinelibrary.wiley.com/doi/abs/10.1002/bbpc.19400460406.
- ↑ History of electron microscopy, 1931–2000. Authors.library.caltech.edu (2002-12-10). Retrieved on 2017-04-29.
- ↑ "North America's first electron microscope". https://news.wsu.edu/2018/05/14/wsu-home-north-americas-first-electron-microscope/.
- ↑ "James Hillier". 2003-05-01. http://web.mit.edu/Invent/iow/hillier.html.
- ↑ Kirkland, Earl (2010). Advanced computing in electron microscopy. New York: Springer. ISBN 978-1-4419-6533-2. OCLC 668095602.
- ↑ "A primer to single-particle cryo-electron microscopy". Cell 161 (3): 438–449. April 2015. doi:10.1016/j.cell.2015.03.050. PMID 25910204.
- ↑ Erni, Rolf; Rossell, MD; Kisielowski, C; Dahmen, U (2009). "Atomic-Resolution Imaging with a Sub-50-pm Electron Probe". Physical Review Letters 102 (9): 096101. doi:10.1103/PhysRevLett.102.096101. PMID 19392535. Bibcode: 2009PhRvL.102i6101E. http://escholarship.org/uc/item/3cs0m4vr.
- ↑ "The Scale of Things". Office of Basic Energy Sciences, U.S. Department of Energy. 2006-05-26. http://www.sc.doe.gov/bes/scale_of_things.html.
- ↑ O'Keefe MA; Allard LF (2004-01-18). Sub-Ångstrom Electron Microscopy for Sub-Ångstrom Nano-Metrology. Information Bridge: DOE Scientific and Technical Information – Sponsored by OSTI. http://www.osti.gov/bridge/servlets/purl/821768-E3YVgN/native/821768.pdf.
- ↑ Burgess, Jeremy (1987). Under the Microscope: A Hidden World Revealed. CUP Archive. p. 11. ISBN 978-0-521-39940-1. https://books.google.com/books?id=30A5AAAAIAAJ&pg=PA11.
- ↑ "Introduction to Electron Microscopy". FEI Company. p. 15. http://www.fei.com/uploadedfiles/documents/content/introduction_to_em_booklet_july_10.pdf.
- ↑ Humbel, Bruno M; Schwarz, Heinz; Tranfield, Erin M; Fleck, Roland A (February 15, 2019). "Chapter 10: Chemical Fixation". Biological Field Emission Scanning Electron Microscopy, First Edition. John Wiley & Sons Ltd.. pp. 191–221. doi:10.1002/9781118663233.ch10. ISBN 9781118663233. https://onlinelibrary.wiley.com/doi/10.1002/9781118663233.ch10.
- ↑ "Cryo-electron microscopy of vitreous sections of native biological cells and tissues". J Struct Biol 148 (1): 131–5. October 2004. doi:10.1016/j.jsb.2004.03.010. PMID 15363793.
- ↑ 31.0 31.1 "Preparing samples from whole cells using focused-ion-beam milling for cryo-electron tomography". Nat Protoc 15 (6): 2041–2070. June 2020. doi:10.1038/s41596-020-0320-x. PMID 32405053.
- ↑ Luft, J.H. (1961). "Improvements in epoxy resin embedding methods". The Journal of Biophysical and Biochemical Cytology 9 (2): p. 409. PMID 13764136.
- ↑ Gruijters, W. T.; Kistler, J; Bullivant, S; Goodenough, D. A. (1987-03-01). "Immunolocalization of MP70 in lens fiber 16-17-nm intercellular junctions". The Journal of Cell Biology 104 (3): 565–572. doi:10.1083/jcb.104.3.565. PMID 3818793.
- ↑ da Silva, Pedro Pinto; Branton, Daniel (1970-06-01). "Membrane splitting in freeze-etching". The Journal of Cell Biology 45 (3): 598–605. doi:10.1083/jcb.45.3.598. PMID 4918216.
- ↑ Rash, J. E.; Johnson, T. J.; Hudson, C. S.; Giddings, F. D.; Graham, W. F.; Eldefrawi, M. E. (1982-11-01). "Labelled-replica techniques: post-shadow labelling of intramembrane particles in freeze-fracture replicas". Journal of Microscopy 128 (Pt 2): 121–138. doi:10.1111/j.1365-2818.1982.tb00444.x. PMID 6184475.
- ↑ Reynolds, E. S. (1963). "The use of lead citrate at high pH as an electron-opaque stain in electron microscopy". Journal of Cell Biology 17 (1): 208–212. doi:10.1083/jcb.17.1.208. PMID 13986422.
- ↑ "Ultrastructure of Golgi apparatus of exocrine cells of mouse pancreas". Exp Cell Res 7 (2): 415–29. November 1954. doi:10.1016/s0014-4827(54)80087-5. PMID 13220587.
- ↑ "Methods in Cell Biology | Correlative Light and Electron Microscopy III | ScienceDirect.com by Elsevier". https://www.sciencedirect.com/bookseries/methods-in-cell-biology/vol/140/.
- ↑ "Chemical approaches to unraveling the biology of mycobacteria". Cell Chem Biol 30 (5): 420–435. May 2023. doi:10.1016/j.chembiol.2023.04.014. PMID 37207631.
- ↑ 40.0 40.1 "Volume electron microscopy". Nat Rev Methods Primers 2: 51. July 2022. doi:10.1038/s43586-022-00131-9. PMID 37409324.
- ↑ "Three dimensional reconstructions of spherical viruses by fourier synthesis from electron micrographs". Nature 226 (5244): 421–5. May 1970. doi:10.1038/226421a0. PMID 4314822. Bibcode: 1970Natur.226..421C.
- ↑ "A practical guide to starting SEM array tomography—An accessible volume EM technique". Chapter 7 - A practical guide to starting SEM array tomography—An accessible volume EM technique. Methods in Cell Biology. 177. 2023. pp. 171–196. doi:10.1016/bs.mcb.2022.12.023. ISBN 9780323916073. https://www.sciencedirect.com/science/article/pii/S0091679X22002114.
- ↑ "Work smart, not hard: How array tomography can help increase the ultrastructure data output". J Microsc. August 2023. doi:10.1111/jmi.13217. PMID 37626455.
- ↑ "Serial block-face scanning electron microscopy to reconstruct three-dimensional tissue nanostructure". PLOS Biol 2 (11): e329. November 2004. doi:10.1371/journal.pbio.0020329. PMID 15514700.
- ↑ "The Mind of a Mouse". Cell 182 (6): 1372–1376. September 2020. doi:10.1016/j.cell.2020.08.010. PMID 32946777.
- ↑ "The functional universe of membrane contact sites". Nat Rev Mol Cell Biol 21 (1): 7–24. January 2020. doi:10.1038/s41580-019-0180-9. PMID 31732717.
- ↑ de Jonge, N.; Ross, F.M. (2011). "Electron microscopy of specimens in liquid". Nature Nanotechnology 6 (8): 695–704. doi:10.1038/nmat944. PMID 12872162. Bibcode: 2003NatMa...2..532W.
- ↑ Gai, P.L.; Boyes, E.D. (2009). "Advances in atomic resolution in situ environmental transmission electron microscopy and 1A aberration corrected in situ electron microscopy". Microsc Res Tech 72 (3): 153–164. doi:10.1002/jemt.20668. PMID 19140163.
- ↑ Adrian, Marc; Dubochet, Jacques; Lepault, Jean; McDowall, Alasdair W. (1984). "Cryo-electron microscopy of viruses". Nature 308 (5954): 32–36. doi:10.1038/308032a0. PMID 6322001. Bibcode: 1984Natur.308...32A. https://serval.unil.ch/notice/serval:BIB_BEC796503260.
- ↑ Sabanay, I.; Arad, T.; Weiner, S.; Geiger, B. (1991). "Study of vitrified, unstained frozen tissue sections by cryoimmunoelectron microscopy". Journal of Cell Science 100 (1): 227–236. doi:10.1242/jcs.100.1.227. PMID 1795028.
- ↑ Kasas, S.; Dumas, G.; Dietler, G.; Catsicas, S.; Adrian, M. (2003). "Vitrification of cryoelectron microscopy specimens revealed by high-speed photographic imaging". Journal of Microscopy 211 (1): 48–53. doi:10.1046/j.1365-2818.2003.01193.x. PMID 12839550.
Library resources about Electron microscopy |
- An Introduction to Microscopy : resources for teachers and students
- Cell Centered Database – Electron microscopy data
- Science Aid: Electron Microscopy:By Kaden park
Original source: https://en.wikipedia.org/wiki/Electron microscope.
Read more |