AFM-IR
AFM-IR (atomic force microscope infrared-spectroscopy) is one of a family of techniques[1][2][3][4][5][6][7][8][9][10][11][12][13][14][15] that are derived from a combination of two parent instrumental techniques; infrared spectroscopy and scanning probe microscopy (SPM). The term was first used to denote a method that combined a tuneable free electron laser with an atomic force microscope (a type of SPM) equipped with a sharp probe that measured the local absorption of infrared light by a sample; it required that the sample be coupled to an infrared-transparent prism and be less than 1μm thick. It improved the spatial resolution of photothermal AFM-based techniques from microns[7] to circa 100 nm.[8][9][10]
Recording the amount of infrared absorption as a function of wavelength or wavenumber creates an infrared absorption spectra that can be used to chemically characterize and even identify unknown materials.[12][15][16] Recording the infrared absorption as a function of position can be used to create chemical composition maps that show the spatial distribution of different chemical components. Novel extensions of the original AFM-IR technique and earlier techniques[1][2][3][4][6][7][16] have enabled the development of bench-top devices capable of nanometer spatial resolution, that do not require a prism and can work with thicker samples, and thereby greatly improving ease of use and expanding the range of samples that can be analysed. One of these techniques has achieved spatial resolutions down to around 20 nm, with a sensitivity down to the scale of molecular monolayer[17]
AFM-IR is related to techniques such as tip-enhanced Raman spectroscopy (TERS), scanning near-field optical microscopy (SNOM),[18] nano-FTIR and other methods of vibrational analysis with scanning probe microscopy.
History
The earliest measurements combining AFM with infrared spectroscopy were performed in 1999 by Hammiche et al. at the University of Lancaster in the United Kingdom,[1] in an EPSRC-funded project led by M Reading and H M Pollock. Separately, Anderson at the Jet Propulsion Laboratory in the United States made a related measurement in 2000.[2] Both groups used a conventional Fourier transform infrared spectrometer (FTIR) equipped with a broadband thermal source, the radiation was focused near the tip of a probe that was in contact with a sample. The Lancaster group obtained spectra by detecting the absorption of infrared radiation using a temperature sensitive thermal probe. Anderson[2] took the different approach of using a conventional AFM probe to detect the thermal expansion. He reported an interferogram but not a spectrum; the first infrared spectrum obtained in this way was reported by Hammiche et al. in 2004:[6] this represented the first proof that spectral information about a sample could be obtained using this approach.
Both of these early experiments used a broadband source in conjunction with an interferometer; these techniques could, therefore, be referred to as AFM-FTIR although Hammiche et al. coined the more general term photothermal microspectroscopy or PTMS in their first paper.[1] PTMS has various subgroups;[19] including techniques that measure temperature[1][3][4][6][7][14][20] measure thermal expansion[2][6][8][9][10][11][12][13] use broadband sources.[1][2][3][4][6][7] use lasers[8][9][10][11][12][12][20] excite the sample using evanescent waves,[8][9][10][11][15] illuminate the sample directly from above[1][2][3][4][5][6][7][12][14][17][20] etc. and different combinations of these. Fundamentally, they all exploit the photothermal effect. Different combinations of sources, methods, methods of detection and methods of illumination have benefits for different applications.[6] Care should be taken to ensure that it is clear which form of PTMS is being used in each case. Currently there is no universally accepted nomenclature. The original technique dubbed AFM-IR that induced resonant motion in the probe using a Free Electron Laser has developed by exploiting the foregoing permutations so that it has evolved into various forms.
The pioneering experiments of Hammiche et al and Anderson had limited spatial resolution due to thermal diffusion - the spreading of heat away from the region where the infrared light was absorbed. The thermal diffusion length (the distance the heat spreads) is inversely proportional to the root of the modulation frequency. Consequently, the spatial resolution achieved by the early AFM-IR approaches was around one micron or more, due to the low modulation frequencies of the incident radiation created by the movement of the mirror in the interferometer. Also, the first thermal probes were Wollaston wire devices[1][2][3][4][5][6] that were developed originally for Microthermal analysis[21] (in fact PTMS was originally considered to be one of a family of microthermal techniques[4]). The comparatively large size of these probes also limited spatial resolution. Bozec et al.[3] and Reading et al.[7] used thermal probes with nanoscale dimensions and demonstrated higher spatial resolution. Ye et al [22] described a MEM-type thermal probe giving sub-100 nm spatial resolution, which they used for nanothermal analysis. The process of exploring laser sources began in 2001 by Hammiche et al when they acquired the first spectrum using a tuneable laser (see Resolution improvement with pulsed laser source).
A significant development was the creation by Reading et al. in 2001[4] of a custom interface that allowed measurements to be made while illuminating the sample from above; this interface focused the infrared beam to a spot of circa 500μm diameter, close to the theoretical maximum[Note 1]. The use of top-down or top-side illumination has the important benefit that samples of arbitrary thickness can be studied on arbitrary substrates. In many cases this can be done without any sample preparation. All subsequent experiments by Hammiche, Pollock, Reading and their co-workers were made using this type of interface including the instrument constructed by Hill et al. for nanoscale imaging using a pulsed laser.[12] The work of the University of Lancaster group in collaboration with workers from the University of East Anglia led to the formation of a company, Anasys Instruments, to exploit this and related technologies[23] (see Commercialization).
Resolution improvement with pulsed laser sources
In the first paper on AFM-based infrared by Hammiche et al.,[1] the relevant well-established theoretical considerations were outlined that predict that high spatial resolution can be achieved using rapid modulation frequencies because of the consequent reduction in the thermal diffusion length. They estimated that spatial resolutions in the range of 20 nm-30 nm should be achievable.[24] The most readily available sources that can achieve high modulation frequencies are pulsed lasers: even when the rapidity of the pulses is not high, the square wave form of a pulse contains very high modulation frequencies in Fourier space. In 2001, Hammiche et al. used a type of bench-top tuneable, pulsed infrared laser known as an optical parametric oscillator or OPO and obtained the first probe-based infrared spectrum with a pulsed laser, however, they did not report any images[16]
Nanoscale spatial resolution AFM-IR imaging using a pulsed laser was first demonstrated by Dazzi et al[8] at the University of Paris-Sud, France. Dazzi and his colleagues used a wavelength-tuneable, free electron laser at the CLIO facility[Note 2] in Orsay, France to provide an infrared source with short pulses. Like earlier workers,[2][6] they used a conventional AFM probe to measure thermal expansion but introduced a novel optical configuration: the sample was mounted on an IR-transparent prism so that it could be excited by an evanescent wave. Absorption of short infrared laser pulses by the sample caused rapid thermal expansion that created a force impulse at the tip of the AFM cantilever. The thermal expansion pulse induced transient resonant oscillations of the AFM cantilever probe. This has led to the technique being dubbed Photo-Thermal Induced Resonance (PTIR), by some workers in the field.[10][16] Some prefer the terms PTIR or PTMS[1][3][5][6][7] to AFM-IR as the technique is not necessarily restricted to infrared wavelengths. The amplitude of the cantilever oscillation is directly related to the amount of infrared radiation absorbed by the sample.[25][26][27][28][29][30][31] By measuring the cantilever oscillation amplitude as a function of wavenumber, Dazzi's group was able to obtain absorption spectra from nanoscale regions of the sample. Compared to earlier work, this approach improved spatial resolution because the use of short laser pulses reduced the duration of the thermal expansion pulse to the point that the thermal diffusion lengths can be on the scale of nanometres rather than microns.
A key advantage of the use of a tuneable laser source, with a narrow wavelength range, is the ability to rapidly map the locations of specific chemical components on the sample surface. To achieve this, Dazzi's group tuned their free electron laser source to a wavelength corresponding to the molecular vibration of the chemical of interest, then mapped the cantilever oscillation amplitude as function of position across the sample. They demonstrated the ability to map chemical composition in E. coli bacteria. They could also visualize polyhydroxybutyrate (PHB) vesicles inside Rhodobacter capsulatus cells[27] and monitor the efficiency of PHB production by the cells.
At the University of East Anglia in the UK, as part of an EPSRC-funded project led by M. Reading and S. Meech, Hill and his co-workers[12] followed the earlier work of Reading et al.[4] and Hammiche et al.[6] and measured thermal expansion using an optical configuration that illuminated the sample from above[5] in contrast to Dazzi et al. who excited the sample with an evanescent wave from below.[8] Hill also made use of an optical parametric oscillator as the infrared source in the manner of Hammiche et al.[16] This novel combination of topside illumination,[4] OPO source[16] and measuring thermal expansion[2][6][8][8] proved capable of nanoscale spatial resolution for infrared imaging and spectroscopy (the figures show a schematic of the UEA apparatus and results obtained with it). The use by Hill and co-workers of illumination from above allowed a substantially wider range of samples to be studied than was possible using Dazzi's technique. By introducing the use of a bench top IR source and topdown illumination, the work of Hammiche, Hill and their coworkers made possible the first commercially viable SPM-based infrared instrument (see Commercialization).
Commercialization
The AFM-IR technique based on a pulsed infrared laser source was commercialized by Anasys Instruments, a company founded by Reading, Hammiche and Pollock in the United Kingdom in 2004;[23][32] a sister, United States corporation was founded a year later. Anasys Instruments developed its product with support from the National Institute of Standards and Technology and the National Science Foundation. Since free electron lasers are rare and available only at select institutions, a key to enabling a commercial AFM-IR was to replace them with a more compact type of infrared source. Following the lead given by Hammiche et al in 2001[16] and Hill et al in 2008,[12] Anasys Instruments introduced an AFM-IR product in early 2010, using a tabletop laser source based on a nanosecond optical parametric oscillator.[28] The OPO source enabled nanoscale infrared spectroscopy over a tuning range of roughly 1000–4000 cm−1 or 2.5-10 μm.
The initial product required samples to be mounted on infrared-transparent prisms, with the infrared light being directed from below in the manner of Dazzi et al.[Note 3] For best operation, this illumination scheme required thin samples, with optimal thickness of less than 1 μm,[16] prepared on the surface of the prism. In 2013, Anasys released an AFM-IR instrument based on the work of Hill et al.[12][20] that supported top-side illumination.
It is worth noting that the first infrared spectrum obtained by measuring thermal expansion using an AFM was obtained by Hammiche and co-workers[6] without inducing resonant motions in the probe cantilever. In this early example the modulation frequency was too low to achieve high spatial resolution but there is nothing, in principle, preventing the measurement of thermal expansion at higher frequencies without analysing or inducing resonant behaviour.[1] Possible options for measuring the displacement of the tip rather than the subsequent propagation of waves along the cantilever include; interferometry focused at the end of the cantilever where the tip is located, a torsional motion resulting from an offset probe (it would only be influenced by the motions of the cantilever as a second order effect) and exploiting the fact that the signal from a heated thermal probe is strongly influenced by the position of the tip relative to the surface thus this could provide a measurement of thermal expansion that wasn’t strongly influenced by or dependent upon resonance. The advantages of a non-resonant method of detection is that any frequency of light modulation could be used thus depth information could be obtained in a controlled way (see below) whereas methods that rely on resonance are limited to harmonics. The thermal-probe based method of Hammiche et al.[1] has found a significant number of applications.[14][20]
A unique application made possible by the top-down illumination combined with a thermal probe[4] is localized depth profiling,[20] this is not possible using either using the Dazzi et al. configuration of AFM-IR or that of Hill et al. despite the fact the latter uses top-down illumination. Obtaining linescans[4][34] and images[20] with thermal probes has been shown to be possible, sub-diffraction limit spatial resolution can be achieved[4] and the resolution for delineating boundaries can be enhanced using chemometric techniques.[20][34]
In all of these examples a spectrum is acquired that spans the entire mid-IR range for each pixel, this is considerably more powerful than measuring the absorption of a single wavelength as is the case for AFM-IR when using either the method of Dazzi et al. or Hill et al. Reading and his group demonstrated how, because thermal probes can be heated, localized thermal analysis[4][20][21] can be combined with photothermal infrared spectroscopy using a single probe. In this way local chemical information could be complemented with local physical properties such melting and glass transition temperatures.[21] This in turn led to the concept of thermally assisted nanosampling,[5][20] where the heated tip performs a local thermal analysis experiment then the probe is retracted taking with it down to femtograms[Note 4] of softened material that adhere to the tip.[30] This material can then be manipulated and/or analysed by photothermal infrared spectroscopy or other techniques.[5][35][36][37][38] This considerably increases the analytical power of this type of SPM-based infrared instrument beyond anything that can be achieved with conventional AFM probes such as those used in AFM-IR when using either the Dazzi et al. or the Hill et al. version.
Thermal probe techniques have still not achieved the nanoscale spatial resolution that thermal expansion methods have attained though this is theoretically possible. For this, a robust thermal probe and a high intensity source is needed. Recently, the first images using a QCL and a thermal probe have been obtained by Reading et al.[39] A good signal to noise ratio enabled rapid imaging but sub-micron spatial resolution was not clearly demonstrated. Theory predicts improvements in spatial resolution could be achieved by confining data analysis to the early part of the thermal response to a step change increase in the intensity of the incident radiation. In this way pollution of the measurement from adjacent regions would be avoided, i.e. the measurement window could be confined to a suitable fraction of the time of flight of the thermal wave (using a Fourier analysis of the response could provide a similar outcome by using the high frequency components). This could be achieved by tapping the probe in synchrony with the laser. Similarly, lasers that provide very rapid modulations could further reduce thermal diffusion lengths.
Although most effort to date has been focused on thermal expansion measurements, this might change. Truly robust thermal probes have recently become available,[40] as have affordable compact QCL’s that are tuneable over a broad frequency range. Consequently, it may soon be the case that thermal probe techniques will become as widely used as those based on thermal expansion. Ultimately, instruments that can easily switch between modes and even combine them using a single probe will certainly become available, for example, a single probe will eventually be able to measure both temperature and thermal expansion.
Recent improvements in AFM-IR
The original commercial AFM-IR instruments required most samples to be thicker than 50 nm to achieve sufficient sensitivity. Sensitivity improvements were achieved using specialized cantilever probes with an internal resonator[[41] and by wavelet based signal processing techniques.[42] Sensitivity was further improved by Lu et al.[17] by using quantum cascade laser (QCL) sources. The high repetition rate of the QCL allows absorbed infrared light to continuously excite the AFM tip at a "contact resonance"[Note 5] of the AFM cantilever. This resonance-enhanced AFM-IR, in combination with electric field enhancement from metallic tips and substrates led to the demonstration of AFM-IR spectroscopy and compositional imaging of films as thin as single self-assembled monolayers.[17]
The AFM-IR has also been integrated with other sources including a picosecond OPO[16] offering a tuning range 1.55 μm to 16 μm (from 6450 cm−1 to 625 cm−1).
Nanospectroscopy
AFM-IR enables nanoscale infrared spectroscopy,[43] i.e. the ability to obtain infrared absorption spectra from nanoscale regions of a sample.
Chemical compositional mapping AFM-IR can also be used to perform chemical imaging or compositional mapping with spatial resolution down to ~20 nm, limited only by the radius of the AFM tip. In this case, the tuneable infrared source emits a single wavelength, corresponding to a specific molecular resonance, i.e. a specific infrared absorption band. By mapping the AFM cantilever oscillation amplitude as a function of position, it is possible to map out the distribution of specific chemical components. Compositional maps can be made at different absorption bands to reveal the distribution of difference chemical species.
AFM-IR nanospectroscopy of a laser printer toner particle, showing spatially resolved chemical analysis. Toner particles are typically complex composites of various binders and transfer agents; these can be revealed by AFM-IR
AFM-IR compositional mapping of Streptomyces bacteria. Left: AFM topographic image of bacterial cells. Middle: AFM-IR absorption at 1650 cm−1, corresponding to the amide I band associated with protein. Right: AFM-IR absorption at the carbonyl band 1740 cm−1, indicating the distribution of triglyceride vesicles within bacterial cells.
Complementary mechanical mapping
The AFM-IR technique can simultaneously provide complementary measurements of the mechanical stiffness and dissipation of a sample surface. When infrared light is absorbed by the sample the resulting rapid thermal expansion excites a "contact resonance" of the AFM cantilever, i.e. a coupled resonance resulting from the properties of both the cantilever and the stiffness and damping of the sample surface. Specifically, the resonance frequency shifts to higher frequencies for stiffer materials and to lower frequencies for softer material. Additionally, the resonance becomes broader for materials with larger dissipation. These contact resonances have been studied extensively by the AFM community (see, for example, atomic force acoustic microscopy). Traditional contact resonance AFM requires an external actuator to excite the cantilever contact resonances. In AFM-IR these contact resonances are automatically excited every time an infrared pulse is absorbed by the sample. So the AFM-IR technique can measure the infrared absorption by the amplitude of the cantilever oscillation response and the mechanical properties of the sample via the contact resonance frequency and quality factor.
Broadband pulsed laser sources
Reading et al. have explored the use of a broadband QCL combined with thermal expansion measurements.[39] Above, the inability of thermal broadband sources to achieve high spatial resolution is discussed (see history). In this case the frequency of modulation is limited by the mirror speed of the interferometer which, in turn, limits the lateral spatial resolution that can be achieved. When using a broadband QCL the resolution is limited not by the mirror speed but by the modulation frequency of the laser pulses (or other waveforms).[1] The benefit of using a broadband source is that an image can be acquired that comprises an entire spectrum or part of a spectrum for each pixel. This is much more powerful than acquiring images bases on a single wavelength. The preliminary results of Reading et al.[39] show that directing a broadband QCL though an interferometer can give an easily detectable response from a conventional AFM probe measuring thermal expansion.
Applications
Applications of AFM-IR have included polymers,[15][28][30][31][44][45] composites, bacteria,[29][46][47][48] cells,[49][50][51][52] biominerals,[53][54] pharmaceutical sciences,[27][55][56] photonics/nanoantennas,[57][58][59][60] fuel cells,[61] fibers,[31][62] skin[63] hair,[64] metal organic frameworks,[65] microdroplets[66] self-assembled monolayers,[17] nanocrystals,[67] and semiconductors.[68]
Polymers blends, composites, multilayer films and fibers AFM-IR has been used to identify and map polymer components in blends,[31] characterize interfaces in composites,[69] and even reverse engineer multilayer films[15] Additionally AFM-IR has been used to study chemical composition in Poly(3][4-ethylenedioxythiophene) (PEDOT) conducting polymers.[45] and vapor infiltration into polyethylene terephthalate PET fibers.[62]
Life sciences
AFM-IR has been used to spectroscopically characterize infection of bacteria by viruses[48] (Bacteriophages), and also the production of polyhydroxybutyrate (PHB) vesicles inside Rhodobacter capsulatus cells[47] and triglycerides[37] in Streptomyces bacteria (for biofuel applications). AFM-IR has also been used to evaluate and map mineral content, crystallinity, collagen maturity and acid phosphate content via ratiometric analysis of various absorption bands in bone[54] AFM-IR has also been used to perform spectroscopy and chemical mapping of structural lipids in human skin[63] and hair[64]
Fuel cells
AFM-IR has been used to study hydrated Nafion membranes used as separators in fuel cells. The measurements revealed the distribution of free and ionically bound water on the Nafion surface.[61]
Photonic nanoantennas
AFM-IR has been used to study the surface plasmon resonance in heavily silicon-doped indium arsenide microparticles.[68] Gold split ring resonators have been studied for use with Surface-Enhanced Infrared Absorption Spectroscopy. In this case AFM-IR was used to measure the local field enhancement of the plasmonics structures (~30X) at 100 nm spatial resolution.[57][69]
Pharmaceutical sciences
AFM-IR has been used to study miscibility and phase separation in drug polymer blends,[55][56] as well as the chemical analysis of nanocrystalline drug particles as small 90 nm across.[27]
Notes
- ↑ Graham Poulter, Research Director for Specac Instruments, "The energy available in an optical instrument is directly related the product of the area A of any point in the optical system, multiplied by the solid angle Ω filled by the beam at that point. This product, AΩ, is known as the étendue (also referred to as the "throughput" or "luminosity") and remains a constant at all points in the system. When focusing a beam down from say a 5mm diameter spot in a typical FTIR to a 0.5mm diameter spot, the area A is decreased by a factor of 100 and, therefore, the solid angle Ω has to be increased by the same factor. When illuminating something on a flat surface from one side there is a physical limitation that means Ω cannot exceed π steradians (it's illuminated from a complete hemisphere). Depending on the solid angle in the original instrument beam, this immediately puts a working limit on the minimum spot size that can be usefully obtained when focusing the beam down". Poulter designed the optics in the interface described by Reading et al.[4]
- ↑ Centre Laser Infrarouge d'Orsay, Orsay Infrared Laser Centre
- ↑ An arrangement is similar to attenuated total reflectance (ATR) schemes used in conventional infrared spectroscopy
- ↑ One femtogram is 10−15 grammes
- ↑ A contact resonance is a vibrational resonance frequency of an AFM cantilever that occurs when the tip of the AFM is in contact with a sample surface. When the QCL is pulsed synchronously with a contact resonance, the detection of the thermal expansion of the sample from infrared absorption is amplified by the quality factor Q of the contact resonance.
References
- ↑ 1.00 1.01 1.02 1.03 1.04 1.05 1.06 1.07 1.08 1.09 1.10 1.11 1.12 Hammiche, A.; Pollock, H. M.; Reading, M.; Claybourn, M.; Turner, P. H.; Jewkes, K. (1999). "Photothermal FT-IR Spectroscopy: A Step Towards FT-IR Microscopy at a Resolution Better Than the Diffraction Limit". Applied Spectroscopy 53 (7): 810–815. doi:10.1366/0003702991947379. Bibcode: 1999ApSpe..53..810H.
- ↑ 2.0 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 Anderson, M. S. (2000). "Infrared Spectroscopy with an Atomic Force Microscope". Applied Spectroscopy 54 (3): 349–352. doi:10.1366/0003702001948538. Bibcode: 2000ApSpe..54..349..
- ↑ 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 Hammiche, A.; Bozec, L.; Conroy, M.; Pollock, H. M.; Mills, G.; Weaver, J. M. R.; Price, D. M.; Reading, M. et al. (2000). "Highly localized thermal, mechanical and spectroscopic characterisation of polymers using miniaturized thermal probes". J. Vac. Sci. Technol. B 18 (3): 1322–1332. doi:10.1116/1.591381. Bibcode: 2000JVSTB..18.1322H. https://semanticscholar.org/paper/34cbd3037c2858b0c8d839b16860f96a03016073.
- ↑ 4.00 4.01 4.02 4.03 4.04 4.05 4.06 4.07 4.08 4.09 4.10 4.11 4.12 4.13 4.14 Reading, M.; Price, D. M.; Grandy, D.B.; Smith, R. M.; Bozec, L.; Conroy, M.; Hammiche, A.; Pollock, H. M. (2001). "Microthermal analysis of polymers: current capabilities and future prospects". Macromol. Symp. 167: 45–62. doi:10.1002/1521-3900(200103)167:1<45::aid-masy45>3.0.co;2-n.
- ↑ 5.0 5.1 5.2 5.3 5.4 5.5 5.6 Reading, M.; Grandy, D.; Hammiche, A.; Bozec, L.; Pollock, H. M. (2002). "Thermally assisted nanosampling and analysis using micro-IR spectroscopy and other analytical technciques". Vibrational Spectroscopy 29 (1): 257–260. doi:10.1016/s0924-2031(01)00185-0.
- ↑ 6.00 6.01 6.02 6.03 6.04 6.05 6.06 6.07 6.08 6.09 6.10 6.11 6.12 6.13 6.14 Hammiche, A.; Bozec, L.; Pollock, H. M.; German, M.; Reading, M. (2004). "Progress in near-field photothermal infrared microspectroscopy". Journal of Microscopy 213 (2): 129–134. doi:10.1111/j.1365-2818.2004.01292.x.
- ↑ 7.0 7.1 7.2 7.3 7.4 7.5 7.6 7.7 Reading, M.; Grandy, D.; Pollock, H. M.; Hammiche, A. (2004). "Micro-thermal analysis using a new high resolution thermal probe". United Kingdom SPM Meeting. Nottingham. https://www.researchgate.net/publication/266266112.
- ↑ 8.0 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8 Dazzi, A.; Prazeres, R.; Glotin, F.; Ortega, J. M. (2005). "Local infrared microspectroscopy with subwavelength spatial resolution with an atomic force microscope tip used as a photothermal sensor". Optics Letters 30 (18): 2388–2390. doi:10.1364/OL.30.002388. PMID 16196328. Bibcode: 2005OptL...30.2388D.
- ↑ 9.0 9.1 9.2 9.3 9.4 Dazzi, A.; Glotin, F.; Ortega, J. M. (September 2006). "Subwavelength infrared spectromicroscopy using an AFM as a local absorption sensor". Infrared Physics and Technology 49 (1–2): 113–121. doi:10.1016/j.infrared.2006.01.009. Bibcode: 2006InPhT..49..113D.
- ↑ 10.0 10.1 10.2 10.3 10.4 10.5 Dazzi, A.; Prazeres, R.; Glotin, F.; Ortega, J. M. (2007). "Analysis of nano-chemical mapping performed by an AFM-based ("AFMIR") acousto-optic technique". Ultramicroscopy 107 (12): 1194–1200. doi:10.1016/j.ultramic.2007.01.018. PMID 17382474.
- ↑ 11.0 11.1 11.2 11.3 Dazzi, A. (2008). Kneipp, J.; Lasch, P.. eds. "Sub-100-Nanometer Infrared Spectroscopy and Imaging Based on a Near-Field Photothermal Technique (PTIR)". Biomedical Vibrational Spectroscopy: 291–312. doi:10.1002/9780470283172.ch13. ISBN 9780470283172.
- ↑ 12.00 12.01 12.02 12.03 12.04 12.05 12.06 12.07 12.08 12.09 12.10 Hill, G.A.; Rice, J.H.; Meech, S.R.; Craig, D.; Kuo, P.; Vodopyanov, K.; Reading, M. (2009). "Submicrometer infrared surface imaging using a scanning probe microscope and an optical parametric oscillator laser". Optics Letters 34 (4): 433. doi:10.1364/OL.34.000431. PMID 19373331. Bibcode: 2009OptL...34..431H. (published online, Feb 2008)
- ↑ 13.0 13.1 Vodopyanov, K.; Hill, G. A.; Rice, J. H.; Meech, S. R.; Craig, D. Q. M.; Reading, M. M.; Dazzi, A.; Kjoller, K. et al. (Fall 2009). "Nano-Spectroscopy in the 2.5-10 Micron Wavelength Range Using Atomic Force Microscope". Frontiers in Optics Laser Science XXV.
- ↑ 14.0 14.1 14.2 14.3 Hammiche, A.; Bozec, L.; German, M.J.; Chalmers, J. M.; Everall, N. J.; Poulter, G.; Reading, M.; Grandy, D. B. et al. (2004). "Mid-infrared micro-spectroscopy of difficult samples using near-field photothermal micro-spectroscopy (PTMS)". Spectroscopy 19 (2): 20–42. with erratum, 19(5), 14 May 2004
- ↑ 15.0 15.1 15.2 15.3 15.4 Eby, T.; Gundusharma, U.; Lo, M.; Sahagian, K.; Marcott, C.; Kjoller, K. (June 13, 2012). "Reverse engineering of polymeric multilayers using AFM-based nanoscale IR spectroscopy and thermal analysis". Spectroscopy Europe 24 (3): 18–21.
- ↑ 16.0 16.1 16.2 16.3 16.4 16.5 16.6 16.7 16.8 Bozec, L.; Hammiche, A.; Pollock, H.M.; Conroy, M.; Everall, N. J.; Turi, L. (2001). "Localized phtothermal infrared spectroscopy using a proximal probe". Journal of Applied Physics 90 (10): 5159. doi:10.1063/1.1403671. Bibcode: 2001JAP....90.5159B.
- ↑ 17.0 17.1 17.2 17.3 17.4 Lu, F.; Jin, M.; Belkin, M.A. (2014). "Tip-enhanced infrared nanospectroscopy via molecular expansion force detection". Nature Photonics 8 (4): 307–312. doi:10.1038/nphoton.2013.373. Bibcode: 2014NaPho...8..307L.
- ↑ H M Pollock; D A Smith (2002). "The use of near-field probes for vibrational spectroscopy and photothermal imaging". in J M Chalmers. Handbook of vibrational spectroscopy vol. 2. pp. 1472–92.
- ↑ F L Martin; H M Pollock (2010). "Microspectroscopy as a tool to discriminate nano-molecular cellular alterations in biomedical research". in J A V Narlikar. Oxford Handbook of Nanoscience and Technology vol. 2. pp. 285–336.
- ↑ 20.0 20.1 20.2 20.3 20.4 20.5 20.6 20.7 20.8 20.9 Dai, X.; Moffat, J. G.; Wood, J.; Reading, M. (April 2012). "Thermal scanning probe microscopy in the development of pharmaceuticals". Advanced Drug Delivery Reviews 64 (5): 449–460. doi:10.1016/j.addr.2011.07.008. PMID 21856345.
- ↑ 21.0 21.1 21.2 Gorbunov, V.V.; Grandy, D.; Reading, M.; Tsukruk, V.V. (2009). "7, Micro and Nanoscale Local Thermal Analysis". Thermal Analysis of Polymers, Fundamantals and Applications. John Wiley and Sons.
- ↑ J Ye; M Reading; N Gotzen; G van Assche (2007). "Scanning Thermal Probe Microscopy:NanoThermal Analysis with Raman Microscopy". Microscopy and Analysis 21 (2): S5–S8.
- ↑ 23.0 23.1 "Impact case study (REF3b)". Research Excellence Framework. http://impact.ref.ac.uk/casestudies2/refservice.svc/GetCaseStudyPDF/43579.
- ↑ H M Pollock (2011). "Towards chemical mapping at sub-micron resolution: near-field spectroscopic delineation of interphase boundaries". Materials Science Forum 662: 1–11. doi:10.4028/www.scientific.net/msf.662.1.
- ↑ Lahiri, B.; Holland, G.; Centrone, A. (October 4, 2012). "Chemical Imaging Beyond the Diffraction Limit: Experimental Validation of the PTIR Technique". Small 9 (3): 439–445. doi:10.1002/smll.201200788. PMID 23034929.
- ↑ Dazzi, A.; Glotin, F.; Carminati, R. (2010). "Theory of infrared nano-spectroscopy by Photo Thermal Induced Resonance". Journal of Applied Physics 107 (12): 124519–124519–7. doi:10.1063/1.3429214. Bibcode: 2010JAP...107l4519D.
- ↑ 27.0 27.1 27.2 27.3 Katzenmeyer, Aksyuk V.; Centrone, A. (2013). "Nanoscale Infrared Spectroscopy: Improving the Spectral Range of the Photothermal Induced Resonance Technique". Analytical Chemistry 85 (4): 1972–1979. doi:10.1021/ac303620y. PMID 23363013.
- ↑ 28.0 28.1 28.2 Felts, J. R.; Kjoller, K.; Lo, M.; Prater, C. B.; King, W. P. (August 31, 2012). "Nanometer-scale infrared spectroscopy of heterogeneous polymer nanostructures fabricated by tip-based nanofabrication". ACS Nano 6 (9): 8015–8021. doi:10.1021/nn302620f. PMID 22928657.
- ↑ 29.0 29.1 Mayet, A.; Deiset-Besseau, A.; Prazeres, R.; Ortega, J. M.; Dazzi, A. (2013). "Analysis of bacterial polyhydroxybutyrate production by multimodal nanoimaging". Biotechnology Advances 31 (3): 369–374. doi:10.1016/j.biotechadv.2012.05.003. PMID 22634017.
- ↑ 30.0 30.1 30.2 Kjoller, K.; Prater, C.; Shetty, R. (November 1, 2010). "Polymer characterization using nanoscale infrared spectroscopy". American Laboratory 42 (11).
- ↑ 31.0 31.1 31.2 31.3 Dazzi; Prater, C. B.; Hu, Q.; Chase, D. B.; Rabolt, J. F.; Marcott, C. (2012). "AFM-IR: combining atomic force microscopy and infrared spectroscopy for nanoscale chemical characterization". Applied Spectroscopy 66 (12): 1365–1384. doi:10.1366/12-06804. PMID 23231899. Bibcode: 2012ApSpe..66.1365D.
- ↑ "Anasys Instruments". https://companycheck.co.uk/company/05138048/ANASYS-INSTRUMENTS-LIMITED/directors-secretaries.
- ↑ Final report EPSRC grant EP/C007751/1 (Report). http://www.mike-reading.com/media/docs/FinalReportforEPSRCGrantEPC007751.pdf.
- ↑ 34.0 34.1 "Towards chemical mapping at sub-micron resolution : near-field spectroscopic delineation of interphase boundaries. Pollock, H.M.". Materials Science Forum (662): 1–11. November 2010.
- ↑ Dai, X.; Moffat, J. G.; Mayes, A.G.; Reading, M.; Craig, D.Q. M.; Belton, P.S.; Grandy, D. B. (2009). "Thermal Probe Based Analytical Microscopy: Thermal Analysis and Photothermal Fourier-Transform Infrared Microspectroscopy Together with Thermally Assisted Nanosampling Coupled with Capillary Electrophoresis". Analytical Chemistry 81 (16): 6612–9. doi:10.1021/ac9004869. PMID 20337375.
- ↑ Harding, L.; Qi, S.; Hill, G.; Reading, M.; Craig, D. Q. M. (May 2008). "The development of microthermal analysis and photothermal microspectroscopy as novel approaches to drug-excipient compatibility studies". International Journal of Pharmaceutics 354 (1–2): 149–157. doi:10.1016/j.ijpharm.2007.11.009. PMID 18162342.; 354(1-2)149-5.
- ↑ 37.0 37.1 Moffat, J. G.; Mayes, A. G.; Belton, P. S.; Craig, D. Q. M.; Reading, M. (2009). "Compositional Analysis of Metal Chelating Materials Using Near-Field Photothermal Fourier Transform Infrared Microspectroscopy". Analytical Chemistry 82 (1): 91–7. doi:10.1021/ac800906t. PMID 19957959.
- ↑ Dai, X.; Belton, P.; DeCogan, D.; Moffat, J. G.; Reading, M. (2011). "Thermally induced movement of micro particles observed on a rough surface: A novel observation and its implications for high throughput analysis and synthesis". Thermochimica Acta 517 (4): 121–125. doi:10.1016/j.tca.2011.01.037.
- ↑ 39.0 39.1 39.2 Reading, M.; Hammiche, A.; Pollock, H.M.; Rankl, C.; Rice, J.; Capponi, S.; Grandy, D.. "Two new scanning probe microscopy techniques for photothermal IR imaging and spectroscopy". Royal Society of Chemistry TAC. Cambridge, UK. https://www.researchgate.net/publication/274703915.30 March-1 April 2015
- ↑ "VertiSense™ Scanning Thermal Microscopy (SThM) Probes". Appnano.com. http://www.appnano.com/vertisense-thermal-probes. Retrieved December 12, 2015.
- ↑ Kjoller, K.; Felts, J. R.; Cook, D.; Prater, C. B.; King, W. P. (2010). "High-sensitivity nanometer-scale infrared spectroscopy using a contact mode microcantilever with an internal resonator paddle". Nanotechnology 21 (18): 185705. doi:10.1088/0957-4484/21/18/185705. PMID 20388971. Bibcode: 2010Nanot..21r5705K. 185705
- ↑ Cho, H.; Felts, J. R.; Yu, M. F.; Bergman, L. A.; Vakakis, A. F.; King, W. P. (2013). "Improved Atomic Force microscope Infrared Spectroscopy for Rapid Nanometer-Scale Chemical Identification". Nanotechnology 24 (44): 444007. doi:10.1088/0957-4484/24/44/444007. PMID 24113150. Bibcode: 2013Nanot..24R4007C.444007
- ↑ Pollock, Hubert M.; Kazarian, Sergei G. (2006) (in en). Microspectroscopy in the Mid-Infrared. 1–26. doi:10.1002/9780470027318.a5609.pub2. ISBN 9780470027318.
- ↑ Marcott, C.; Lo, M.; Kjoller, K.; Prater, C.; Noda, I. (2011). "Spatial Differentiation of Sub-Micrometer Domains in a Poly(hydroxyalkanoate) Copolymer Using Instrumentation that Combines Atomic Force Microscopy (AFM) and Infrared (IR) Spectroscopy". Applied Spectroscopy 65 (10): 1145–1150. doi:10.1366/11-06341. PMID 21986074. Bibcode: 2011ApSpe..65.1145M.
- ↑ 45.0 45.1 Ghosh, S.; Remita, H.; Ramos, L.; Dazzi, A.; Deiset-Besseau, A.; Beaunier, P.; Goubard, F.; Aubert, P. H. et al. (2014). "PEDOT nanostructures synthesized in hexagonal mesophases". New Journal of Chemistry 38 (3): 1106–1115. doi:10.1039/c3nj01349a. https://semanticscholar.org/paper/5bdec42a3f8c9612bcb6d4c2c29eac43940615d1.
- ↑ Deiset-Besseau, A.; Prater, C. B.; Virolle, M.J.; Dazzi, A. (2014). "Monitoring TriAcylGlycerols Accumulation by Atomic Force Microscopy Based Infrared Spectroscopy in Streptomyces Species for Biodiesel Applications". The Journal of Physical Chemistry Letters 5 (4): 654–658. doi:10.1021/jz402393a. PMID 26270832.
- ↑ 47.0 47.1 Mayet, A.; Dazzi, A.; Prazeres, R.; Ortega, J. M.; Jaillard, D. (2010). "In situ identification and imaging of bacterial polymer nanogranules by infrared nanospectroscopy". Analyst 135 (10): 2540–2545. doi:10.1039/c0an00290a. PMID 20820491. Bibcode: 2010Ana...135.2540M.
- ↑ 48.0 48.1 Dazzi, A.; Prazeres, R.; Glotin, F.; Ortega, J. M.; Al-Sawaftah, M.; de Frutos, M. (2008). "Chemical mapping of the distribution of viruses into infected bacteria with a photothermal method". Ultramicroscopy 108 (7): 635–641. doi:10.1016/j.ultramic.2007.10.008. PMID 18037564.
- ↑ Clede, S.; Lambert, F.; Sandt, C.; Kascakova, S.; Unger, M.; Harte, E.M.; Plamont, A.; Saint-Fort, R. et al. (2013). "Detection of an estrogen derivative in two breast cancer cell lines using a single core multimodal probe for imaging (SCoMPI) imaged by a panel of luminescent and vibrational techniques". Analyst 138 (19): 5627–5638. doi:10.1039/c3an00807j. PMID 23897394. Bibcode: 2013Ana...138.5627C. https://hal.archives-ouvertes.fr/hal-00849792/file/Cl%C3%A8de_2013_Detection_of_an.pdf.
- ↑ Policar, C.; Waern, J. B.; Plamont, M. A.; Clède, S.; Mayet, C.; Prazeres, R.; Ortega, J. M.; Vessières, A. et al. (2011). "Subcellular IR Imaging of a Metal–Carbonyl Moiety Using Photothermally Induced Resonance". Angewandte Chemie International Edition 123 (4): 890–894. doi:10.1002/ange.201003161.
- ↑ Dazzi, A.; Policar, C. (2011). Chabal, C. M. P. J.. ed. Biointerface Characterization by Advanced IR Spectroscopy. Elsevier, Amsterdam. pp. 245–278.
- ↑ Mayet, C.; Dazzi, A.; Prazeres, R.; Allot, F.; Glotin, F.; Ortega, J. M. (2008). "Sub-100nm IR spectromicroscopy of living cells". Optics Letters 33 (14): 1611–1613. doi:10.1364/OL.33.001611. PMID 18628814. Bibcode: 2008OptL...33.1611M.
- ↑ Marcott, C.; Lo, M.; Hu, Q.; Kjoller, K.; Boskey, A.; Noda, I. (2014). "Using 2D correlation analysis to enhance spectral information available from highly spatially resolved AFM-IR spectra". Journal of Molecular Structure 1069: 284–289. doi:10.1016/j.molstruc.2014.01.036. PMID 25024505. Bibcode: 2014JMoSt1069..284M.
- ↑ 54.0 54.1 Gourio-Arsiquaud, S; Marcott, C.; Hu, Q.; Boskey, A. (2014). "Studying variations in bone composition at nano-scale resolution: a preliminary report". Calcified Tissue International 95 (5): 413–418. doi:10.1007/s00223-014-9909-9. PMID 25155443.
- ↑ 55.0 55.1 Van Eerdenbrugh, B.; Lo, M.; Kjoller, K.; Marcott, C.; Taylor, L. S. (2012). "Nanoscale mid-infrared imaging of phase separation in a drug–polymer blend". Journal of Pharmaceutical Sciences 101 (6): 2066–2073. doi:10.1002/jps.23099. PMID 22388948.
- ↑ 56.0 56.1 Van Eerdenbrugh, B.; Lo, M.; Kjoller, K.; Marcott, C.; Taylor, L. S. (2012). "Nanoscale Mid-Infrared Evaluation of the Miscibility Behavior of Blends of Dextran or Maltodextrin with Poly(vinylpyrrolidone)". Molecular Pharmaceutics 9 (5): 1459–1469. doi:10.1021/mp300059z. PMID 22483035.
- ↑ 57.0 57.1 Lahiri, B.; Holland, G.; Aksyuk, V.; Centrone, A. (2013). "Nanoscale imaging of plasmonic hot spots and dark modes with the photothermal-induced resonance technique". Nano Letters 13 (7): 3218–3224. doi:10.1021/nl401284m. PMID 23777547. Bibcode: 2013NanoL..13.3218L.
- ↑ Felts, J. R.; Law, S.; Roberts, C. M.; Podolskiy, V.; Wasserman, D. M.; King, W. P. (2013). "Near-field infrared absorption of plasmonic semiconductor microparticles studied using atomic force microscope infrared spectroscopy". Applied Physics Letters 102 (15): 152110. doi:10.1063/1.4802211. Bibcode: 2013ApPhL.102o2110F.
- ↑ Katzenmeyer, A. M.; Chae, J.; Kasica, R.; Holland, G.; Lahiri, B.; Centrone, A. (2014). "Nanoscale Imaging and Spectroscopy of Plasmonic Modes with the PTIR Technique". Advanced Optical Materials 2 (8): 718–722. doi:10.1002/adom.201400005. https://zenodo.org/record/1229074.
- ↑ Houel, J.; Sauvage, S.; Boucaud, P.; Dazzi, A.; Prazeres, R.; Glotin, F.; Ortega, J. M.; Miard, A. et al. (2007). "Ultraweak-Absorption Microscopy of a Single Semiconductor Quantum Dot in the Midinfrared Range". Physical Review Letters 99 (21): 217404. doi:10.1103/PhysRevLett.99.217404. PMID 18233255. Bibcode: 2007PhRvL..99u7404H.217404
- ↑ 61.0 61.1 Awatani, T.; Midorikawa, H.; Kojima, N.; Ye, J.; Marcott, C. (2013). "Morphology of water transport channels and hydrophobic clusters in Nafion from high spatial resolution AFM-IR spectroscopy and imaging". Electrochemistry Communications 30: 5–8. doi:10.1016/j.elecom.2013.01.021.
- ↑ 62.0 62.1 Akyildiz, H. I.; Lo, M.; Dillon, E.; Roberts, A. T.; Everitt, H. O.; Jur, J. S. (2014). "Formation of novel photoluminescent hybrid materials by sequential vapor infiltration into polyethylene terephthalate fibers". Journal of Materials Research 29 (23): 2817–2826. doi:10.1557/jmr.2014.333. Bibcode: 2014JMatR..29.2817A.
- ↑ 63.0 63.1 Marcott, C.; Lo, M.; Kjoller, K.; Domanov, Y.; Balooch, G.; Luengo, G. S. (2013). "Nanoscale infrared (IR) spectroscopy and imaging of structural lipids in human stratum corneum using an atomic force microscope to directly detect absorbed light from a tunable IR laser source". Experimental Dermatology 22 (6): 419–421. doi:10.1111/exd.12144. PMID 23651342.
- ↑ 64.0 64.1 Marcott, C.; Lo, M.; Kjoller, K.; Fiat, F.; Baghdadli, N.; Balooch, G.; Luengo, G. S. (2014). "Localization of Human Hair Structural Lipids Using Nanoscale Infrared Spectroscopy and Imaging". Applied Spectroscopy 68 (=5): 564–569. doi:10.1366/13-07328. PMID 25014600. Bibcode: 2014ApSpe..68..564M.
- ↑ Katzenmeyer, A. M.; Canivet, J.; Holland, G.; Farrusseng, D.; Centrone, A. (2014). "Assessing Chemical Heterogeneity at the Nanoscale in Mixed-Ligand Metal–Organic Frameworks with the PTIR Technique". Angewandte Chemie International Edition 53 (11): 2852–2856. doi:10.1002/anie.201309295. PMID 24615798.
- ↑ Muller, T.; Ruggeri, F. S.; Kulik, A. J.; Shimanovich, U.; Mason, T. O.; Knowles, T. P. J.; Dietler, G. (2014). "Nanoscale spatially resolved infrared spectra from single microdroplets". Lab on a Chip 14 (7): 1315–1319. doi:10.1039/C3LC51219C. PMID 24519414.
- ↑ Rosen, E. L.; Buonsanti, R.; Llordes, A.; Sawvel, A. M.; Milliron, D. J.; Helms, B. A. (2012). "Exceptionally Mild Reactive Stripping of Native Ligands from Nanocrystal Surfaces by Using Meerwein's Salt". Angewandte Chemie International Edition 51 (3): 684–689. doi:10.1002/anie.201105996. PMID 22147424.
- ↑ 68.0 68.1 Houel, J.; Homeyer, E.; Sauvage, S.; Boucaud, P.; Dazzi, A.; Prazeres, R.; Ortega, J. M. (2009). "Midinfrared absorption measured at a λ/400 resolution with an atomic force microscope". Opt Express 17 (13): 10887–10894. doi:10.1364/OE.17.010887. PMID 19550489. Bibcode: 2009OExpr..1710887H.
- ↑ 69.0 69.1 "Nanoscale Infrared Spectroscopy of Polymer Composites", americanlaboratory.com
External links
- Infrared Imaging beyond the Diffraction Limit (NIST Andrea Centrone Group) [1]
- Sub-wavelength resolution microspectroscopy (University of Texas Mikhail Belkin group) [2]